PLATINUM China: a prospective, randomized investigation of the platinum chromium everolimus-eluting stent in de novo coronary artery lesions.
The PLATINUM China clinical trial evaluated the safety and effectiveness of the thin-strut, everolimus-eluting, platinum-chromium PROMUS™ Element™ stent (PtCr-EES) (Boston Scientific, Marlborough, MA) for the treatment of patients in China.
Clinical outcomes from the PtCr-EES have not been evaluated in patients from China, nor has it been directly compared with the second-generation stainless steel paclitaxel-eluting TAXUS Liberté stent (PES) in a randomized head-to-head trial. Methods In this prospective, multicenter, single-blind, superiority trial, patients with a single de novo atherosclerotic coronary artery lesion were randomized 1:3 to receive either the PES or PtCr-EES. The primary endpoint was in-stent late loss at 9 months.
Among 127 PES and 373 PtCr-EES patients (71.2% male; mean age 57.3 years), the primary endpoint of 9-month in-stent late loss was 0.40 ± 0.45 mm for PES versus 0.11 ± 0.36 mm for PtCr-EES (P < 0.001). In-stent % diameter stenosis was 22.20 ± 16.00% for PES versus 11.06 ± 13.86% for PtCr-EES (P < 0.001) at 9 months. The 1-year rate of death/MI was 1.6% (2/127) for PES versus 0% (0/371) for PtCr-EES (P = 0.06) and target vessel revascularization was 4.7% (6/127) for PES versus 2.7% (10/371) for PtCr-EES (P = 0.26). No stent thromboses occurred at 12 months in either group.
In the largest prospective angiographic evaluation conducted to date with this stent, the PROMUS Element PtCr-EES was superior to the TAXUS Liberté PES for the primary endpoint of late loss at 9 months, with low rates of clinical events at 1 year. Clinical follow-up will continue to 2 years.
Gao R
,Han Y
,Yang Y
,Zhang J
,Hou Y
,Wang H
,Li H
,Fang Q
,Yu B
,Xu B
,Allocco DJ
,Dawkins KD
... -
《-》
-
Long-Term Safety and Efficacy of Platinum Chromium Everolimus-Eluting Stents in Coronary Artery Disease: 5-Year Results From the PLATINUM Trial.
The authors sought to evaluate the final 5-year safety and effectiveness of the platinum-chromium everolimus-eluting stent (PtCr-EES) in the randomized trial, as well as in 2 single-arm substudies that evaluated PtCr-EES in small vessels (diameter <2.5 mm; n = 94) and long lesions (24 to 34 mm; n = 102).
In the multicenter, randomized PLATINUM (PLATINUM Clinical Trial to Assess the PROMUS Element Stent System for Treatment of De Novo Coronary Artery Lesions), the PtCr-EES was noninferior to the cobalt-chromium everolimus-eluting stent (CoCr-EES) at 1 year in 1,530 patients undergoing percutaneous coronary intervention.
Patients with 1 or 2 de novo coronary artery lesions (reference vessel diameter 2.50 to 4.25 mm, length ≤24 mm) were randomized 1:1 to PtCr-EES versus CoCr-EES. All patients in the substudies received PtCr-EES. The primary endpoint was target lesion failure (TLF), a composite of target vessel-related cardiac death, target vessel-related myocardial infarction, or ischemia-driven target lesion revascularization.
In the randomized trial, the 5-year TLF rate was 9.1% for PtCr-EES and 9.3% for CoCr-EES (hazard ratio [HR]: 0.97; p = 0.87). Landmark analysis demonstrated similar TLF rates from discharge to 1 year (HR: 1.12; p = 0.70) and from 1 to 5 years (HR: 0.90; p = 0.63). There were no significant differences in the rates of cardiac death, myocardial infarction, target lesion or vessel revascularization, or stent thrombosis. PtCr-EES had 5-year TLF rates of 7.0% in small vessels and 13.6% in long lesions.
PtCr-EES demonstrated comparable safety and effectiveness to CoCr-EES through 5 years of follow-up, with low rates of stent thrombosis and other adverse events. The 5-year event rates were also acceptable in patients with small vessels and long lesions treated with PtCr-EES. (The PLATINUM Clinical Trial to Assess the PROMUS Element Stent System for Treatment of De Novo Coronary Artery Lesions [PLATINUM]; NCT00823212; The PLATINUM Clinical Trial to Assess the PROMUS Element Stent System for Treatment of De Novo Coronary Artery Lesions in Small Vessels [PLATINUM SV]; NCT01498692; The PLATINUM Clinical Trial to Assess the PROMUS Element Stent System for Treatment of Long De Novo Coronary Artery Lesions [PLATINUM LL]; NCT01500434).
Kelly CR
,Teirstein PS
,Meredith IT
,Farah B
,Dubois CL
,Feldman RL
,Dens J
,Hagiwara N
,Rabinowitz A
,Carrié D
,Pompili V
,Bouchard A
,Saito S
,Allocco DJ
,Dawkins KD
,Stone GW
... -
《-》
-
A clinical and angiographic study of the XIENCE V everolimus-eluting coronary stent system in the treatment of patients with multivessel coronary artery disease: the EXECUTIVE trial (EXecutive RCT: evaluating XIENCE V in a multi vessel disease).
This study sought to investigate the efficacy and performance of the XIENCE V everolimus-eluting stent (EES) (Abbott Vascular, Santa Clara, California) in the treatment of de novo coronary lesions in patients with 2- to 3-vessel multivessel coronary artery disease (MV-CAD).
Drug-eluting stents (DES) have emerged as an alternative to conventional coronary artery bypass surgery in patients with MV-CAD although first-generation DES yielded inferior efficacy and safety compared with surgery.
Prospective, randomized (1:1), multicenter feasibility trial was designed to assess angiographic efficacy of EES compared with the TAXUS paclitaxel-eluting stent (PES) in 200 patients, and a prospective, open-label, single-arm, controlled registry was designed to analyze the clinical outcome of EES at 1-year follow-up in 400 MV-CAD patients. For the randomized trial, the primary endpoint was in-stent late loss at 9 months. For the registry, the primary endpoint was a composite of all-cause death, myocardial infarction, and ischemia-driven target vessel revascularization at 12 months.
The primary endpoint per single lesion was significantly lower in the EES group compared with the PES group (-0.03 ± 0.49 mm vs. 0.23 ± 0.51 mm, p = 0.001). Similar results were observed when analyzing all lesions (0.05 ± 0.51 mm vs. 0.24 ± 0.50 mm, p < 0.001). Clinical outcome at 1 year yielded a composite of major adverse cardiac events of 9.2% in the single-arm registry, and 11.1% and 16.5% in the EES and PES randomized groups, respectively (p = 0.30).
The EXECUTIVE trial was a randomized pilot trial dedicated to the comparison of the efficacy of 2 different DES among patients with 2- to 3-vessel MV-CAD. The study shows lower in-stent late loss at 9 months with the EES XIENCE V compared with the PES TAXUS Libertè, and a low major adverse cardiac event rate at 1 year in patients with 2-to 3-vessel MV-CAD. (EXECUTIVE [EXecutive RCT: Evaluating XIENCE V in a Multi Vessel Disease]; NCT00531011).
Ribichini F
,Romano M
,Rosiello R
,La Vecchia L
,Cabianca E
,Caramanno G
,Milazzo D
,Loschiavo P
,Rigattieri S
,Musarò S
,Pironi B
,Fiscella A
,Amico F
,Indolfi C
,Spaccarotella C
,Bartorelli A
,Trabattoni D
,Della Rovere F
,Rolandi A
,Beqaraj F
,Belli R
,Sangiorgio P
,Villani R
,Berni A
,Sheiban I
,Lopera Quijada MJ
,Cappi B
,Ribaldi L
,Vassanelli C
,EXECUTIVE Trial Investigators
... -
《-》
-
5-year results of a randomized comparison of XIENCE V everolimus-eluting and TAXUS paclitaxel-eluting stents: final results from the SPIRIT III trial (clinical evaluation of the XIENCE V everolimus eluting coronary stent system in the treatment of patient
This study sought to evaluate the long-term safety and efficacy of everolimus-eluting stents (EES) and paclitaxel-eluting stents (PES) in patients with obstructive coronary artery disease.
The use of EES compared to PES has been shown to result in improved clinical outcomes in patients undergoing PCI. However, there have been concerns regarding the durability of these benefits over longer-term follow-up.
SPIRIT III was a prospective, multicenter trial in which 1,002 patients were randomized 2:1 to EES versus PES. Endpoints included ischemia-driven target vessel failure (TVF) (death, myocardial infarction (MI), or ischemia-driven target vessel revascularization [TVR]), the pre-specified primary endpoint), target lesion failure (TLF) (cardiac death, target-vessel MI, or ischemia-driven target lesion revascularization [TLR]), major adverse cardiac events (MACE) (cardiac death, MI, or ischemia-driven TLR), their individual components and stent thrombosis.
Five-year follow-up was available in 91.9% of patients. Treatment with EES versus PES resulted in lower 5-year Kaplan-Meier rates of TVF (19.3% vs. 24.5%, p = 0.05), TLF (12.7% vs. 19.0%, p = 0.008), and MACE (13.2% vs. 20.7%, p = 0.007). EES also resulted in reduced rates of all-cause death (5.9% vs. 10.1%, p = 0.02), with nonsignificantly different rates of MI, stent thrombosis, and TLR, and no evidence of late catch-up of TLR over time.
At 5 years after treatment, EES compared to PES resulted in durable benefits in composite safety and efficacy measures as well as all-cause mortality. Additionally, the absolute difference in TLR between devices remained stable over time without deterioration of effect during late follow-up.
Gada H
,Kirtane AJ
,Newman W
,Sanz M
,Hermiller JB
,Mahaffey KW
,Cutlip DE
,Sudhir K
,Hou L
,Koo K
,Stone GW
... -
《-》
-
Meta-analysis of everolimus-eluting versus paclitaxel-eluting stents in coronary artery disease: final 3-year results of the SPIRIT clinical trials program (Clinical Evaluation of the Xience V Everolimus Eluting Coronary Stent System in the Treatment of P
This study sought to investigate whether the everolimus-eluting stent (EES) is superior to the paclitaxel-eluting stent (PES) with respect to long-term individual clinical outcomes.
Individual studies have indicated a clinical advantage of coronary EES compared with PES with respect to restenosis and the composite endpoint of major adverse cardiac events. However, these trials were not powered for superiority in low-frequency event rates and have reported limited data beyond 1-year follow-up.
We conducted a meta-analysis of the final 3-year results from the international SPIRIT (Clinical Evaluation of the Xience V Everolimus Eluting Coronary Stent System in the Treatment of Patients With De Novo Native Coronary Artery Lesions) II, III, and IV clinical trials. Individual patient data from 4,989 patients who were prospectively randomized to treatment with EES (n = 3,350) or PES (n = 1,639) were pooled for analysis.
At 3-year follow-up, EES was superior to PES in reducing the following event rates: target lesion failure (8.9% vs. 12.5%, hazard ratio [HR]: 0.71, 95% confidence interval [CI]: 0.59 to 0.85; p = 0.0002), all-cause mortality (3.2% vs 5.1%, HR: 0.65, 95% CI: 0.49 to 0.86; p = 0.003), myocardial infarction (3.2% vs. 5.1%, HR: 0.64, 95% CI: 0.48 to 0.85; p = 0.002), cardiac death or myocardial infarction (4.4% vs. 6.3%, HR: 0.70, 95% CI: 0.54 to 0.90; p = 0.005), ischemia-driven target lesion revascularization (6.0% vs. 8.2%, HR: 0.72, 95% CI: 0.58 to 0.90; p = 0.004), stent thrombosis (0.7% vs. 1.7%, HR: 0.45, 95% CI: 0.26 to 0.78; p = 0.003), and major adverse cardiac events (9.4% vs. 13.0%, HR: 0.71, 95% CI: 0.60 to 0.85; p = 0.0002). No interaction was present between stent type and the 3-year relative rates of target lesion failure across a broad range of subgroups, with the exception of diabetes and vessel (left anterior descending vs. other).
In this large dataset with 3-year follow-up, coronary implantation of EES compared with PES resulted in reduced rates of all-cause mortality, myocardial infarction, ischemia-driven target lesion revascularization, stent thrombosis, and target lesion failure. Further research is warranted to characterize possible interactions between stent type, diabetes, and vessel.
Dangas GD
,Serruys PW
,Kereiakes DJ
,Hermiller J
,Rizvi A
,Newman W
,Sudhir K
,Smith RS Jr
,Cao S
,Theodoropoulos K
,Cutlip DE
,Lansky AJ
,Stone GW
... -
《-》