Long-Term Safety and Efficacy of Platinum Chromium Everolimus-Eluting Stents in Coronary Artery Disease: 5-Year Results From the PLATINUM Trial.
The authors sought to evaluate the final 5-year safety and effectiveness of the platinum-chromium everolimus-eluting stent (PtCr-EES) in the randomized trial, as well as in 2 single-arm substudies that evaluated PtCr-EES in small vessels (diameter <2.5 mm; n = 94) and long lesions (24 to 34 mm; n = 102).
In the multicenter, randomized PLATINUM (PLATINUM Clinical Trial to Assess the PROMUS Element Stent System for Treatment of De Novo Coronary Artery Lesions), the PtCr-EES was noninferior to the cobalt-chromium everolimus-eluting stent (CoCr-EES) at 1 year in 1,530 patients undergoing percutaneous coronary intervention.
Patients with 1 or 2 de novo coronary artery lesions (reference vessel diameter 2.50 to 4.25 mm, length ≤24 mm) were randomized 1:1 to PtCr-EES versus CoCr-EES. All patients in the substudies received PtCr-EES. The primary endpoint was target lesion failure (TLF), a composite of target vessel-related cardiac death, target vessel-related myocardial infarction, or ischemia-driven target lesion revascularization.
In the randomized trial, the 5-year TLF rate was 9.1% for PtCr-EES and 9.3% for CoCr-EES (hazard ratio [HR]: 0.97; p = 0.87). Landmark analysis demonstrated similar TLF rates from discharge to 1 year (HR: 1.12; p = 0.70) and from 1 to 5 years (HR: 0.90; p = 0.63). There were no significant differences in the rates of cardiac death, myocardial infarction, target lesion or vessel revascularization, or stent thrombosis. PtCr-EES had 5-year TLF rates of 7.0% in small vessels and 13.6% in long lesions.
PtCr-EES demonstrated comparable safety and effectiveness to CoCr-EES through 5 years of follow-up, with low rates of stent thrombosis and other adverse events. The 5-year event rates were also acceptable in patients with small vessels and long lesions treated with PtCr-EES. (The PLATINUM Clinical Trial to Assess the PROMUS Element Stent System for Treatment of De Novo Coronary Artery Lesions [PLATINUM]; NCT00823212; The PLATINUM Clinical Trial to Assess the PROMUS Element Stent System for Treatment of De Novo Coronary Artery Lesions in Small Vessels [PLATINUM SV]; NCT01498692; The PLATINUM Clinical Trial to Assess the PROMUS Element Stent System for Treatment of Long De Novo Coronary Artery Lesions [PLATINUM LL]; NCT01500434).
Kelly CR
,Teirstein PS
,Meredith IT
,Farah B
,Dubois CL
,Feldman RL
,Dens J
,Hagiwara N
,Rabinowitz A
,Carrié D
,Pompili V
,Bouchard A
,Saito S
,Allocco DJ
,Dawkins KD
,Stone GW
... -
《-》
-
PLATINUM China: a prospective, randomized investigation of the platinum chromium everolimus-eluting stent in de novo coronary artery lesions.
The PLATINUM China clinical trial evaluated the safety and effectiveness of the thin-strut, everolimus-eluting, platinum-chromium PROMUS™ Element™ stent (PtCr-EES) (Boston Scientific, Marlborough, MA) for the treatment of patients in China.
Clinical outcomes from the PtCr-EES have not been evaluated in patients from China, nor has it been directly compared with the second-generation stainless steel paclitaxel-eluting TAXUS Liberté stent (PES) in a randomized head-to-head trial. Methods In this prospective, multicenter, single-blind, superiority trial, patients with a single de novo atherosclerotic coronary artery lesion were randomized 1:3 to receive either the PES or PtCr-EES. The primary endpoint was in-stent late loss at 9 months.
Among 127 PES and 373 PtCr-EES patients (71.2% male; mean age 57.3 years), the primary endpoint of 9-month in-stent late loss was 0.40 ± 0.45 mm for PES versus 0.11 ± 0.36 mm for PtCr-EES (P < 0.001). In-stent % diameter stenosis was 22.20 ± 16.00% for PES versus 11.06 ± 13.86% for PtCr-EES (P < 0.001) at 9 months. The 1-year rate of death/MI was 1.6% (2/127) for PES versus 0% (0/371) for PtCr-EES (P = 0.06) and target vessel revascularization was 4.7% (6/127) for PES versus 2.7% (10/371) for PtCr-EES (P = 0.26). No stent thromboses occurred at 12 months in either group.
In the largest prospective angiographic evaluation conducted to date with this stent, the PROMUS Element PtCr-EES was superior to the TAXUS Liberté PES for the primary endpoint of late loss at 9 months, with low rates of clinical events at 1 year. Clinical follow-up will continue to 2 years.
Gao R
,Han Y
,Yang Y
,Zhang J
,Hou Y
,Wang H
,Li H
,Fang Q
,Yu B
,Xu B
,Allocco DJ
,Dawkins KD
... -
《-》
-
Meta-analysis of everolimus-eluting versus paclitaxel-eluting stents in coronary artery disease: final 3-year results of the SPIRIT clinical trials program (Clinical Evaluation of the Xience V Everolimus Eluting Coronary Stent System in the Treatment of P
This study sought to investigate whether the everolimus-eluting stent (EES) is superior to the paclitaxel-eluting stent (PES) with respect to long-term individual clinical outcomes.
Individual studies have indicated a clinical advantage of coronary EES compared with PES with respect to restenosis and the composite endpoint of major adverse cardiac events. However, these trials were not powered for superiority in low-frequency event rates and have reported limited data beyond 1-year follow-up.
We conducted a meta-analysis of the final 3-year results from the international SPIRIT (Clinical Evaluation of the Xience V Everolimus Eluting Coronary Stent System in the Treatment of Patients With De Novo Native Coronary Artery Lesions) II, III, and IV clinical trials. Individual patient data from 4,989 patients who were prospectively randomized to treatment with EES (n = 3,350) or PES (n = 1,639) were pooled for analysis.
At 3-year follow-up, EES was superior to PES in reducing the following event rates: target lesion failure (8.9% vs. 12.5%, hazard ratio [HR]: 0.71, 95% confidence interval [CI]: 0.59 to 0.85; p = 0.0002), all-cause mortality (3.2% vs 5.1%, HR: 0.65, 95% CI: 0.49 to 0.86; p = 0.003), myocardial infarction (3.2% vs. 5.1%, HR: 0.64, 95% CI: 0.48 to 0.85; p = 0.002), cardiac death or myocardial infarction (4.4% vs. 6.3%, HR: 0.70, 95% CI: 0.54 to 0.90; p = 0.005), ischemia-driven target lesion revascularization (6.0% vs. 8.2%, HR: 0.72, 95% CI: 0.58 to 0.90; p = 0.004), stent thrombosis (0.7% vs. 1.7%, HR: 0.45, 95% CI: 0.26 to 0.78; p = 0.003), and major adverse cardiac events (9.4% vs. 13.0%, HR: 0.71, 95% CI: 0.60 to 0.85; p = 0.0002). No interaction was present between stent type and the 3-year relative rates of target lesion failure across a broad range of subgroups, with the exception of diabetes and vessel (left anterior descending vs. other).
In this large dataset with 3-year follow-up, coronary implantation of EES compared with PES resulted in reduced rates of all-cause mortality, myocardial infarction, ischemia-driven target lesion revascularization, stent thrombosis, and target lesion failure. Further research is warranted to characterize possible interactions between stent type, diabetes, and vessel.
Dangas GD
,Serruys PW
,Kereiakes DJ
,Hermiller J
,Rizvi A
,Newman W
,Sudhir K
,Smith RS Jr
,Cao S
,Theodoropoulos K
,Cutlip DE
,Lansky AJ
,Stone GW
... -
《-》
-
A randomized comparison of platinum chromium-based everolimus-eluting stents versus cobalt chromium-based Zotarolimus-Eluting stents in all-comers receiving percutaneous coronary intervention: HOST-ASSURE (harmonizing optimal strategy for treatment of cor
This study sought to test whether the newly developed platinum chromium (PtCr)-based everolimus-eluting stent (EES) is noninferior to the cobalt chromium (CoCr)-based zotarolimus-eluting stent (ZES) in all-comers receiving percutaneous coronary intervention (PCI).
PtCr provides improved radial strength, conformability, and visibility compared with the CoCr alloy, but PtCr-based stents have not been tested in a wide range of patients receiving PCI. Also, recent case series have raised the issue of longitudinal stent deformation (LSD) with newer drug-eluting stents.
We randomly assigned 3,755 all-comers receiving PCI to PtCr-EES or CoCr-ZES. The primary outcome was target lesion failure (TLF) at 1-year post-PCI, defined as the composite of cardiac death, nonfatal target vessel-related myocardial infarction, and ischemia-driven target lesion revascularization. Post-hoc angiographic analysis was performed to qualitatively and quantitatively analyze LSD.
At 1 year, TLF occurred in 2.9% and 2.9% of the population in the PtCr-EES and CoCr-ZES groups, respectively (superiority p = 0.98, noninferiority p = 0.0247). There were no significant differences in the individual components of TLF as well as the patient-oriented clinical outcome. Of 5,010 stents analyzed, LSD occurred in 0.2% and 0% in the PtCr-EES and CoCr-ZES groups, respectively (p = 0.104). There was no significant difference in post-deployment stent length ratio between the 2 stents (p = 0.352).
At 1 year, PtCr-EES was noninferior to CoCr-ZES in all-comers receiving PCI. Although LSD was observed only in PtCr-EES, both the stent length ratio and the frequency of LSD were not significantly different between the 2 stent types, and PtCr-EES was not associated with adverse clinical outcomes. (Harmonizing Optimal Strategy for Treatment of Coronary Artery Stenosis-SAfety & EffectiveneSS of Drug-ElUting Stents & Anti-platelet REgimen [HOST-ASSURE]; NCT01267734).
Park KW
,Kang SH
,Kang HJ
,Koo BK
,Park BE
,Cha KS
,Rhew JY
,Jeon HK
,Shin ES
,Oh JH
,Jeong MH
,Kim S
,Hwang KK
,Yoon JH
,Lee SY
,Park TH
,Moon KW
,Kwon HM
,Hur SH
,Ryu JK
,Lee BR
,Park YW
,Chae IH
,Kim HS
,HOST–ASSURE Investigators
... -
《-》
-
Small-vessel PCI outcomes in men, women, and minorities following platinum chromium everolimus-eluting stents: Insights from the pooled PLATINUM Diversity and PROMUS Element Plus Post-Approval studies.
We evaluated 1-year outcomes after platinum chromium everolimus-eluting stents (PtCr-EES) in small versus non-small coronary arteries within a large, diverse sample of men, women, and minorities.
There exists limited outcomes data on the use of second-generation drug-eluting stent to treat small diameter coronary arteries.
We pooled patients from the PLATINUM Diversity and PROMUS Element Plus stent registries. Small-vessel percutaneous coronary intervention (SV-PCI) was defined as ≥1 target lesion with reference vessel diameter (RVD) ≤2.5 mm. Endpoints included major adverse cardiac event (MACE; death, myocardial infarction [MI] or target vessel revascularization [TVR]), target vessel failure (TVF; death related to the target vessel, target vessel MI or TVR) and definite/probable stent thrombosis (ST). Multivariable Cox regression was used to risk-adjust outcomes.
We included 4,155/4,182 (99%) patients with available RVD, of which 1,607 (39%) underwent small-vessel PCI. SV-PCI was not associated with increased MACE (adjHR 1.02; 95%CI 0.81-1.30) or TVF (adjHR 1.07; 95%CI 0.82-1.39). MI risk was lower in white men compared to women and minorities, both in the setting of SV-PCI (adjHR 0.41; 95%CI 0.23-0.74 and adjHR 0.39; 95%CI 0.20-0.75, respectively) and for non-SV-PCI (adjHR 0.61; 95%CI 0.38-0.99 and adjHR 0.45; 95%CI 0.27-0.74, respectively). There was no significant interaction between RVD and sex or minority status for any endpoint.
In a large diverse contemporary PCI outcomes database, SV-PCI with PtCr-EES was not associated with increased MACE or TVR and did not account for the increased MI risk noted in women and minorities compared to white men.
Guedeney P
,Claessen BE
,Mehran R
,Kandzari DE
,Aquino M
,Davis S
,Tamis L
,Wang JC
,Othman I
,Gigliotti OS
,Haghighat A
,Singh S
,Lopez M
,Giugliano G
,Horwitz PA
,Sorrentino S
,Underwood P
,Allocco D
,Meredith IT
,Batchelor W
... -
《-》