Microdosimetric measurements of a monoenergetic and modulated Bragg Peaks of 62 MeV therapeutic proton beam with a synthetic single crystal diamond microdosimeter.
摘要:
The purpose of this study was to investigate for the first time the performance of a synthetic single crystal diamond detector for the microdosimetric characterization of clinical 62 MeV ocular therapy proton beams. A novel diamond microdosimeter with a well-defined sensitive volume was fabricated and tested with a monoenergetic and spread-out Bragg peak (SOBP) of the CATANA therapeutic proton beam in Catania, Italy. The whole sensitive volume of the detector has an active planar-sectional area of 100 µm × 100 µm and a thickness of approximately 6.3 um. Microdosimetric measurements were performed at several water equivalent depths, corresponding to positions of clinical relevance. From the measured spectra, microdosimetric quantities such as the frequency mean lineal energy ( y ¯ F ), dose mean lineal energy ( y ¯ D ) as well as microdosimetric relative biological effectiveness (RBEµ ) values were derived for each depth along both a pristine Bragg curve and SOBP. Finally, Geant4 Monte Carlo simulations were performed modeling the detector geometry and CATANA beamline in order to calculate the average linear energy transfer (LET) values in the diamond active layer and water. The microdosimetric spectra acquired by the diamond microdosimeter show different shapes as a function of the water equivalent depths. No spectral distortion, due to pile-up events and polarization effects, was observed. The experimental spectra have a very low detection threshold due to the electronic noise during the irradiation of about 1 keV/μm. The y ¯ F and y ¯ D values were in agreement with expected trends, showing a sharp increase in mean lineal energy at the distal edge of the Bragg peak. In addition, a good agreement between the mean lineal energy values and the calculated average LET ones was also observed. Finally, the RBE values evaluated with the diamond microdosimeter were in excellent agreement with those obtained with a mini tissue equivalent proportional counter as well as with radiobiological measurements in the same proton beam field. The microdosimetric performance of the tested synthetic single crystal diamond microdosimeter clearly indicates its suitability for quality assurance in clinical proton therapy beam.
收起
展开
DOI:
10.1002/mp.14466
被引量:
年份:
1970


通过 文献互助 平台发起求助,成功后即可免费获取论文全文。
求助方法1:
知识发现用户
每天可免费求助50篇
求助方法1:
关注微信公众号
每天可免费求助2篇
求助方法2:
完成求助需要支付5财富值
您目前有 1000 财富值
相似文献(359)
参考文献(0)
引证文献(1)
来源期刊
影响因子:暂无数据
JCR分区: 暂无
中科院分区:暂无