A novel, yet simple MLC-based 3D-crossfire technique for spatially fractionated GRID therapy treatment of deep-seated bulky tumors.
Treating deep-seated bulky tumors with traditional single-field Cerrobend GRID-blocks has many limitations such as suboptimal target coverage and excessive skin toxicity. Heavy traditional GRID-blocks are a concern for patient safety at various gantry-angles and dosimetric detail is not always available without a GRID template in user's treatment planning system. Herein, we propose a simple, yet clinically useful multileaf collimator (MLC)-based three-dimensional (3D)-crossfire technique to provide sufficient target coverage, reduce skin dose, and potentially escalate tumor dose to deep-seated bulky tumors.
Thirteen patients (multiple sites) who underwent conventional single-field cerrobend GRID-block therapy (maximum, 15 Gy in 1 fraction) were re-planned using an MLC-based 3D-crossfire method. Gross tumor volume (GTV) was used to generate a lattice pattern of 10 mm diameter and 20 mm center-to-center mimicking conventional GRID-block using an in-house MATLAB program. For the same prescription, MLC-based 3D-crossfire grid plans were generated using 6-gantry positions (clockwise) at 60° spacing (210°, 270°, 330°, 30°, 90°, 150°, therefore, each gantry angle associated with a complement angle at 180° apart) with differentially-weighted 6 or 18 MV beams in Eclipse. For each gantry, standard Millenium120 (Varian) 5 mm MLC leaves were fit to the grid-pattern with 90° collimator rotation, so that the tunneling dose distribution was achieved. Acuros-based dose was calculated for heterogeneity corrections. Dosimetric parameters evaluated include: mean GTV dose, GTV dose heterogeneities (peak-to-valley dose ratio, PVDR), skin dose and dose to other adjacent critical structures. Additionally, planning time and delivery efficiency was recorded. With 3D-MLC, dose escalation up to 23 Gy was simulated for all patient's plans.
All 3D-MLC crossfire GRID plans exhibited excellent target coverage with mean GTV dose of 13.4 ± 0.5 Gy (range: 12.43-14.24 Gy) and mean PVDR of 2.0 ± 0.3 (range: 1.7-2.4). Maximal and dose to 5 cc of skin were 9.7 ± 2.7 Gy (range: 5.4-14.0 Gy) and 6.3 ± 1.8 Gy (range: 4.1-11.1 Gy), on average respectively. Three-dimensional-MLC treatment planning time was about an hour or less. Compared to traditional GRID-block, average beam on time was 20% less, while providing similar overall treatment time. With 3D-MLC plans, tumor dose can be escalated up to 23 Gy while respecting skin dose tolerances.
The simple MLC-based 3D-crossfire GRID-therapy technique resulted in enhanced target coverage for de-bulking deep-seated bulky tumors, reduced skin toxicity and spare adjacent critical structures. This simple MLC-based approach can be easily adopted by any radiotherapy center. It provides detailed dosimetry and a safe and effective treatment by eliminating the heavy physical GRID-block and could potentially provide same day treatment. Prospective clinical trial with higher tumor-dose to bulky deep-seated tumors is anticipated.
Pokhrel D
,Halfman M
,Sanford L
,Chen Q
,Kudrimoti M
... -
《Journal of Applied Clinical Medical Physics》
-
Conebeam CT-guided 3D MLC-based spatially fractionated radiation therapy for bulky masses.
For fast, safe, and effective management of large and bulky (≥8 cm) non-resectable tumors, we have developed a conebeam CT-guided three-dimensional (3D)-conformal MLC-based spatially fractionated radiation therapy (SFRT) treatment. Using an in-house MLC-fitting algorithm, Millennium 120 leaves were fitted to the gross tumor volume (GTV) generating 1-cm diameter holes at 2-cm center-to-center distance at isocenter. SFRT plans of 15 Gy were generated using four to six coplanar crossfire gantry angles 60° apart with a 90° collimator, differentially weighted with 6- or 10-MV beams. A dose was calculated using AcurosXB algorithm, generating sieve-like dose channels without post-processing the physician-drawn GTV contour within an hour of CT simulation allowing for the same day treatment. In total, 50 extracranial patients have been planned and treated using this method, comprising multiple treatment sites. This novel MLC-fitting algorithm provided excellent dose parameters with mean GTV (V7.5 Gy) and mean GTV doses of 53.2% and 7.9 Gy, respectively, for 15 Gy plans. Average peak-to-valley dose ratio was 3.2. Mean beam-on time was 3.32 min, and treatment time, including patient setup and CBCT to beam-off, was within 15 min. Average 3D couch correction from original skin-markers was <1.0 cm. 3D MLC-based SFRT plans enhanced target dose for bulky masses, including deep-seated large tumors while protecting skin and adjacent critical organs. Additionally, it provides the same day, safe, effective, and convenient treatment by eliminating the risk to therapists and patients from heavy gantry-mounted physical GRID-block-we recommend other centers to use this simple and clinically useful method. This rapid SFRT planning technique is easily adoptable in any radiation oncology clinic by eliminating the need for plan optimization and patient-specific quality assurance times while providing dosimetry information in the treatment planning system. This potentially allows for dose-escalation to deep-seated masses to debulk unresectable large tumors providing an option for neoadjuvant treatment. An outcome study of clinical trial is underway.
Pokhrel D
,Bernard ME
,Mallory R
,St Clair W
,Kudrimoti M
... -
《Journal of Applied Clinical Medical Physics》
-
Comparison of treatment planning approaches for spatially fractionated irradiation of deep tumors.
The purpose of this work was to compare the dosimetry and delivery times of 3D-conformal (3DCRT)-, volumetric modulated arc therapy (VMAT)-, and tomotherapy-based approaches for spatially fractionated radiation therapy for deep tumor targets.
Two virtual GRID phantoms were created consisting of 7 "target" cylinders (1-cm diameter) aligned longitudinally along the tumor in a honey-comb pattern, mimicking a conventional GRID block, with 2-cm center-to-center spacing (GRID2 cm ) and 3-cm center-to-center spacing (GRID3 cm ), all contained within a larger cylinder (8 and 10 cm in diameter for the GRID2 cm and GRID3 cm , respectively). In a single patient, a GRID3 cm structure was created within the gross tumor volume (GTV). Tomotherapy, VMAT (6 MV + 6 MV-flattening-filter-free) and multi-leaf collimator segment 3DCRT (6 MV) plans were created using commercially available software. Two tomotherapy plans were created with field widths (TOMO2.5 cm ) 2.5 cm and (TOMO5 cm ) 5 cm. Prescriptions for all plans were set to deliver a mean dose of 15 Gy to the GRID targets in one fraction. The mean dose to the GRID target and the heterogeneity of the dose distribution (peak-to-valley and peak-to-edge dose ratios) inside the GRID target were obtained. The volume of normal tissue receiving 7.5 Gy was determined.
The peak-to-valley ratios for GRID2 cm /GRID3 cm /Patient were 2.1/2.3/2.8, 1.7/1.5/2.8, 1.7/1.9/2.4, and 1.8/2.0/2.8 for the 3DCRT, VMAT, TOMO5 cm , and TOMO2.5 cm plans, respectively. The peak-to-edge ratios for GRID2 cm /GRID3 cm /Patient were 2.8/3.2/5.4, 2.1/1.8/5.4, 2.0/2.2/3.9, 2.1/2.7/5.2 and for the 3DCRT, VMAT, TOMO5 cm , and TOMO2.5 cm plans, respectively. The volume of normal tissue receiving 7.5 Gy was lowest in the TOMO2.5 cm plan (GRID2 cm /GRID3 cm /Patient = 54 cm3 /19 cm3 /10 cm3 ). The VMAT plans had the lowest delivery times (GRID2 cm /GRID3 cm /Patient = 17 min/8 min/9 min).
Our results present, for the first time, preliminary evidence comparing IMRT-GRID approaches which result in high-dose "islands" within a target, mimicking what is achieved with a conventional GRID block but without high-dose "tail" regions outside of the target. These approaches differ modestly in their ability to achieve high peak-to-edge ratios and also differ in delivery times.
Sheikh K
,Hrinivich WT
,Bell LA
,Moore JA
,Laub W
,Viswanathan AN
,Yan Y
,McNutt TR
,Meyer J
... -
《Journal of Applied Clinical Medical Physics》
-
Dosimetric impact of intrafraction patient motion on MLC-based 3D-conformal spatially fractionated radiation therapy treatment of large and bulky tumors.
To evaluate the dosimetric impact on spatially fractionated radiation therapy (SFRT) plan quality due to intrafraction patient motion via multi-field MLC-based method for treating large and bulky (≥8 cm) unresectable tumors.
For large tumors, a cone beam CT-guided 3D conformal MLC-based SFRT method was utilized with 15 Gy prescription. An MLC GTV-fitting algorithm provided 1 cm diameter apertures with a 2 cm center-to-center distance at the isocenter. This generated a highly heterogeneous sieve-like dose distribution within an hour, enabling same-day SFRT treatment. Fifteen previously treated SFRT patients were analyzed (5 head & neck [H&N], 5 chest and lungs, and 5 abdominal and pelvis masses). For each plan, intrafraction motion errors were simulated by incrementally shifting original isocenters of each field in different x-, y-, and z-directions from 1 to 5 mm. The dosimetric metrics analyzed were: peak-to-valley-dose-ratio (PVDR), percentage of GTV receiving 7.5 Gy, GTV mean dose, and maximum dose to organs-at-risk (OARs).
For ±1, ±2, ±3, ±4, and ±5 mm isocenter shifts: PVDR dropped by 3.9%, 3.8%, 4.0%, 4.1%, and 5.5% on average respectively. The GTV(V7.5) remained within 0.2%, and the GTV mean dose remained within 3.3% on average, compared to the original plans. The average PVDR drop for 5 mm shifts was 4.2% for H&N cases, 10% for chest and lung, and 2.2% for abdominal and pelvis cases. OAR doses also increased. The maximum dose to the spinal cord increased by up to 17 cGy in H&N plans, mean lung dose (MLD) changed was small for chest/lung, but the bowel dose varied up to 100 cGy for abdominal and pelvis cases.
Due to tumor size, location, and characteristics of MLC-based SFRT, isocenter shifts of up to ±5 mm in different directions had moderate effects on PVDR for H&N and pelvic tumors and a larger effect on chest tumors. The dosimetric impact on OAR doses depended on the treatment site. Site-specific patient masks, Vac-Lok bags, and proper immobilization devices similar to SBRT/SRT setups should be used to minimize these effects.
Misa J
,Volk A
,Bernard ME
,Clair WS
,Pokhrel D
... -
《Journal of Applied Clinical Medical Physics》
-
A treatment planning approach to spatially fractionated megavoltage grid therapy for bulky lung cancer.
The purpose of this study was to explore the treatment planning methods of spatially fractionated megavoltage grid therapy for treating bulky lung tumors using multileaf collimator (MLC). A total of 5 patients with lung cancer who had gross tumor volumes ranging from 277 to 635 cm(3) were retrospectively chosen for this study. The tumors were from 6.5 to 9.6 cm at shortest dimension. Several techniques using either electronic compensation or intensity-modulated radiation therapy (IMRT) were used to create a variety of grid therapy plans on the Eclipse treatment planning system. The dose prescription point was calculated to the volume, and a dose of 20 Gy with 6-MV/15-MV beams was used in each plan. The dose-volume histogram (DVH) curves were obtained to evaluate dosimetric characteristics. In addition, DVH curves from a commercially available cerrobend grid collimator were also used for comparison. The linear-quadratic radiobiological response model was used to assess therapeutic ratios (TRs) and equivalent uniform doses (EUD) for all generated plans. A total of 6 different grid therapy plans were created for each patient. Overall, 4 plans had different electronic compensation techniques: Ecomps-Tubes, Ecomps-Circles, Ecomps-Squares, and Ecomps-Weave; the other 2 plans used IMRT and IMRT-Weave techniques. The DVH curves and TRs demonstrated that these MLC-based grid therapy plans can achieve dosimetric properties very similar to those of the cerrobend grid collimator. However, the MLC-based plans have larger EUDs than those with the cerrobend grid collimator. In addition, the field shaping can be performed for targets of any shape in MLC-based plans. Thus, they can deliver a more conformal dose to the targets and spare normal structures better than the cerrobend grid collimator can. The plans generated by the MLC technique demonstrated the advantage over the standard cerrobend grid collimator on accommodating targets and sparing normal structures. Overall, 6 different plans showed 6 different dosimetric parameters. However, an optimal grid therapy plan selection from among these 6 types requires more information from clinical trials and radiobiological studies.