Radiation therapy of synchronous bilateral breast carcinoma (SBBC) using multiple techniques.
The purpose of this study was to establish intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) treatment plans for synchronous bilateral breast cancer (SBBC) and to compare those plans with the previous treatment plans using 3D conformal radiation therapy (3DCRT). The differences among the treatments were also statistically compared regarding dosimetry distribution and treatment efficiency. The research was conducted with 10 SBBC patients. The study established IMRT (12 fields with a single isocenter) and VMAT (2 partial arcs with a single isocenter) treatment plans for SBBC patients and then compared those plans with 3DCRT (8 fields with multiple isocenters). The plans were evaluated based on a dose-volume histogram analysis. For planning target volumes (PTVs), the mean doses and the values of V95%, V105%, conformity index, and homogeneity index were reported. For the organs at risk, the analysis included the mean dose, maximum dose, and VXGy, depending on the organs (lungs, heart, and liver). To objectively evaluate the efficiency of the treatment plans, each plan's beam times, treatment times (including set-up time), and monitor units were compared. Tukey test and one-way analysis of variance were used to compare the PTV and organs at risk values of the 3 techniques. Additionally, the independent-samples t-test was used to compare the 2 techniques (IMRT and VMAT) based on the values of Rt. PTV and Lt. PTV (p < 0.05). For PTV dose distribution, IMRT showed increases of approximately 1.2% in Dmean and of approximately 5.7% in V95% dose distribution compared with 3DCRT. In comparison to VMAT, 3DCRT showed about 3.0% higher dose distribution in Dmean and V95%. IMRT was the best in terms of conformity index and homogeneity index (p < 0.05), whereas 3DCRT and VMAT did not significantly differ from each other. In terms of dose distribution on lungs, heart, and liver, the percentage of volume at high doses such as V30Gy and V40Gy was approximately 70% lower for IMRT and approximately 40% lower for VMAT than for 3DCRT. For distribution volumes of low doses such as V5% and V10%, that for 3DCRT was approximately 60% smaller than for IMRT and approximately 70% smaller than for VMAT. Comparison between IMRT and VMAT showed that the IMRT was superior in all distribution factors. VMAT showed better treatment efficiency than 3DCRT or IMRT. Among the SBBC radiotherapy treatment plans, IMRT was superior to 3DCRT and VMAT in terms of PTV dose distribution, whereas VMAT showed the most outstanding treatment efficiency.
Kim SJ
,Lee MJ
,Youn SM
《-》
Evaluation of fixed-jaw IMRT and tangential partial-VMAT radiotherapy plans for synchronous bilateral breast cancer irradiation based on a dosimetric study.
To investigate the fixed-jaw intensity-modulated radiotherapy (F-IMRT) and tangential partial volumetric modulated arc therapy (tP-VMAT) treatment plans for synchronous bilateral breast cancer (SBBC).
Twelve SBBC patients with pTis-2N0M0 stages who underwent whole-breast irradiation after breast-conserving surgery were planned with F-IMRT and tP-VMAT techniques prescribing 42.56 Gy (2.66 Gy*16f) to the breast. The F-IMRT used 8-12 jaw-fixed tangential fields with single (sF-IMRT) or two (F-IMRT) isocenters located under the sternum or in the center of the left and right planning target volumes (PTVs), and tP-VMAT used 4 tangential partial arcs with two isocenters located in the center of the left and right PTVs. Plan evaluation was based on dose-volume histogram (DVH) analysis. Dosimetric parameters were calculated to evaluate plan quality; total monitor units (MUs), and the gamma analysis for patient-specific quality assurance (QA) were also evaluated.
For PTVs, the three plans had similar Dmean and conformity index (CI) values. F-IMRT showed a slightly better target coverage according to the V100% values and demonstrated an obvious reduction in V105% and Dmax compared with the values observed for sF-IMRT and tP-VMAT. Compared with tP-VMAT, sF-IMRT was slightly better in terms of V100% , V105% and Dmax . In addition, F-IMRT achieved the best homogeneity index (HI) values for PTVs. Concerning healthy tissue, tP-VMAT had an advantage in minimizing the high dose volume. The MUs of the tP-VMAT plan were decreased approximately 1.45 and 1 times compared with the sF-IMRT and F-IMRT plans, respectively, and all plans passed QA. For the lungs, heart and liver, F-IMRT achieved the smallest values in terms of Dmean and showed a significant difference compared with tP-VMAT. Simultaneously, sF-IMRT was also superior to tP-VMAT. For the coronary artery, tP-VMAT achieved the lowest Dmean , while the value for F-IMRT was 2.24% lower compared with sF-IMRT. For all organs at risk (OARs), tP-VMAT was superior at the high dose level. In contrast, sF-IMRT and F-IMRT were obviously superior at the low dose level. The sF-IMRT and F-IMRT plans showed consistent trends.
All treatment plans for the provided techniques were of high quality and feasible for SBBC patients. However, we recommend F-IMRT with a single isocenter as a priority technique because of the tremendous advantage of local hot spot control in PTVs and the reduced dose to OARs at low dose levels. When the irradiated dose to the lungs and heart exceed the clinical restriction, two isocenter F-IMRT can be used to maximize OAR sparing. Additionally, tP-VMAT can be adopted for improving cold spots in PTVs or high-dose exposure to normal tissue when the interval between PTVs is narrow.
Huang JH
,Wu XX
,Lin X
,Shi JT
,Ma YJ
,Duan S
,Huang XB
... -
《Journal of Applied Clinical Medical Physics》
Dosimetric comparison of helical tomotherapy, volumetric-modulated arc therapy, intensity-modulated radiotherapy, and field-in-field technique for synchronous bilateral breast cancer.
To compare the dosimetric characteristics of helical tomotherapy (HT), volumetric-modulated arc therapy (VMAT), intensity-modulated radiotherapy (IMRT), and tangential field-in-field technique (FIF) for the treatment of synchronous bilateral breast cancer (SBBC).
Ten patients with early-stage unilateral breast cancer were selected for simulating the patients with SBBC in this retrospective analysis. Treatment plans with HT, VMAT, IMRT, and FIF were generated for each patient with a total dose of 50.4 Gy in 28 fractions to the target. Plan quality, namely conformity index (CI), homogeneity index (HI), dose-volume statistics of organs at risk (OARs), and beam-on time (BOT), were evaluated.
HT plans showed a lower mean heart dose (3.53 ± 0.31Gy) compared with the other plans (VMAT = 5.6 ± 1.36 Gy, IMRT = 3.80 ± 0.76 Gy, and FIF = 4.84 ± 2.13 Gy). Moreover, HT plans showed a significantly lower mean lung dose (p < 0.01) compared with the other plans: mean right lung doses were 6.81 ± 0.67, 10.32 ± 1.04, 9.07 ± 1.21, and 10.03 ± 1.22 Gy and mean left lung doses were 6.33 ± 0.87, 8.82 ± 0.91, 7.84 ± 1.07, and 8.64 ± 0.99 Gy for HT, VMAT, IMRT, and FIF plans, respectively. The mean dose to the left anterior descending artery was significantly lower in HT plans (p < 0.01) than in the other plans: HT = 19.41 ± 0.51 Gy, VMAT = 25.77 ± 7.23 Gy, IMRT = 27.87 ± 6.48 Gy, and FIF = 30.95 ± 10.17 Gy. FIF plans showed a worse CI and HI compared with the other plans. VMAT plans showed shorter BOT (average, 3.9 ± 0.2 minutes) than did HT (average, 11.0 ± 3.0 minutes), IMRT (average, 6.1 ± 0.5 minutes), and FIF (average, 4.6 ± 0.7 minutes) plans.
In a dosimetric comparison for SBBC, HT provided the most favorable dose sparing of OARs. However, HT with longer BOT may increase patient discomfort and treatment uncertainty. VMAT enabled shorter BOT with acceptable doses to OARs and had a better CI than did FIF and IMRT.
Cheng HW
,Chang CC
,Shiau AC
,Wang MH
,Tsai JT
... -
《-》
Dosimetric evaluation of conventional radiotherapy, 3-D conformal radiotherapy and direct machine parameter optimisation intensity-modulated radiotherapy for breast cancer after conservative surgery.
The use of conservative surgery combined with whole-breast irradiation (WBI) has been established as a valid alternative to mastectomy for the management of early-stage breast cancer. The aim of this study was to compare dosimetric parameters of the planning target volume(PTV) and organs at risk (OARs) between conventional radiation therapy (CR), 3-D conformal radiation therapy (3DCRT), and direct machine parameter optimisation intensity-modulated radiation therapy (DMPO-IMRT) after breast-conserving surgery.
Computed tomography (CT) scans from 20 patients (13 left-sided and 7 right-sided) previously treated with T1N0 or ductal carcinoma were selected for this dosimetric planning study. We designed CR, 3DCRT and DMPO-IMRT plans for each patient. The prescribed dose was 50 Gy/2 Gy/25 f, 95% of PTV received the prescription dose. Doses were computed with a commercially available treatment planning system using convolution/superimposition (CS) algorithm. Plans were compared according to dose-volume histogram (DVH) analysis in terms of PTV homogeneity and conformity indices (HI and CI) as well as OARs dose and volume parameters.
Both the HI and CI of the PTV showed statistically significant difference between CR, 3DCRT and DMPO-IMRT with those of DMPO-IMRT were best (P < 0.05). Compared with CR, 3DCRT showed smaller exposed volumes of ipsilateral lung, contralateral breast and heart while DMPO-IMRT indicated larger exposed volumes of ipsilateral lung (except for V20 and V30), contralateral breast and heart. In addition, DMPO-IMRT demonstrated an increase of exposed volume of ipsilateral lung (except for V30), contralateral breast and heart compared with 3DCRT.
In WBI of breast cancer after conservative surgery, 3DCRT and DMPO-IMRT improved the homogeneity and conformity of the PTV compared with CR. Meanwhile, 3DCRT reduced the irradiated volumes of OARs at all dose levels listed in our study while DMPO-IMRT reduced the irradiated volumes of OARs in high-dose areas but increased the irradiated volumes of OARs in low-dose areas.
Zhang F
,Zheng M
《-》