Optimizing finger systolic blood pressure measurements with laser Doppler: Validation of the second phalanx site.
摘要:
Finger systolic blood pressure measurement (FSBP) has been shown helpful in the detection of distal arterial insufficiency in upper limbs. This work assesses the possibility to measure FSBP on the 2nd phalanx instead of the first one in order to improve its sensitivity and to verify this would not alter the repeatability of the measurement. In this multicenter study, FSBP was measured twice in all fingers but the thumbs in consecutive systemic sclerosis patients on the first phalanx and the second phalanx in alternate order using laser-Doppler flowmetry. Thirty-seven patients were enrolled. The repeatability of FSBP was excellent and similar on the first and 2nd phalanxes with coefficients of variation respectively of 7.1% and 7.6%. While the correlation between the FSBP at the two sites was fair (Pearson coefficient 0.69; p < 0.001). The agreement was poor, with a mean difference of 14 mm Hg between the two sites. Significantly higher differences were found in fingers with digital ulcers. The ROC curves showed a better prediction of the 2nd phalanx measurements. FSBP has an excellent repeatability whatever the site of phalanx. However, measurements performed on the 2nd phalanx have a better sensitivity for the prediction of digital ulcers.
收起
展开
DOI:
10.1016/j.mvr.2020.104029
被引量:
年份:
1970


通过 文献互助 平台发起求助,成功后即可免费获取论文全文。
求助方法1:
知识发现用户
每天可免费求助50篇
求助方法1:
关注微信公众号
每天可免费求助2篇
求助方法2:
完成求助需要支付5财富值
您目前有 1000 财富值
相似文献(103)
参考文献(0)
引证文献(1)
来源期刊
影响因子:暂无数据
JCR分区: 暂无
中科院分区:暂无