Thin Composite-Wire-Strut Zotarolimus-Eluting Stents Versus Ultrathin-Strut Sirolimus-Eluting Stents in BIONYX at 2 Years.
The aim of this study was to assess 2-year safety and efficacy of the current-generation thin composite-wire-strut durable-polymer Resolute Onyx zotarolimus-eluting stent (ZES), compared with the ultrathin-strut biodegradable-polymer Orsiro sirolimus-eluting stent (SES) in all-comers and a pre-specified small-vessel subgroup analysis.
The Resolute Onyx ZES is widely used in clinical practice, but no follow-up data beyond 1 year have been published. The randomized BIONYX (Bioresorbable Polymer-Coated Orsiro Versus Durable Polymer-Coated Resolute Onyx Stents) trial (NCT02508714) established the noninferiority of ZES versus SES regarding target vessel failure (TVF) rates.
A total of 2,488 all-comer patients were treated at 7 coronary intervention centers in Belgium, Israel, and the Netherlands. The main endpoint, TVF, was a composite of safety (cardiac death or target vessel-related myocardial infarction) and efficacy (clinically indicated target vessel revascularization). Two-year follow-up data were analyzed using Kaplan-Meier methods.
Two-year follow-up data were available for 2,460 of 2,488 patients (98.9%). TVF occurred in 93 of 1,243 patients (7.6%) assigned to ZES versus 87 of 1,245 patients (7.1%) assigned to SES (log-rank p = 0.66). There was no significant between-stent difference in individual components of this endpoint. The incidence of definite-or-probable stent thrombosis was low for both treatment arms (0.4% vs. 1.1%; log-rank p = 0.057). In patients stented in small vessels, there was no between-stent difference (TVF 8.2% vs. 8.7% [log-rank p = 0.75], target lesion revascularization 4.0% vs. 4.4% [log-rank p = 0.77]).
At 2-year follow-up, the novel thin composite-wire-strut durable-polymer Resolute Onyx ZES showed in all-comers similar safety and efficacy compared with the ultrathin cobalt-chromium-strut biodegradable-polymer Orsiro SES. The analysis of patients who were treated in small vessels also suggested no advantage for either stent.
Buiten RA
,Ploumen EH
,Zocca P
,Doggen CJM
,Jessurun GAJ
,Schotborgh CE
,Roguin A
,Danse PW
,Benit E
,Aminian A
,van Houwelingen KG
,Schramm AR
,Stoel MG
,Somi S
,Hartmann M
,Linssen GCM
,von Birgelen C
... -
《-》
-
Thin, Very Thin, or Ultrathin Strut Biodegradable or Durable Polymer-Coated Drug-Eluting Stents: 3-Year Outcomes of BIO-RESORT.
The aim of this study was to assess the 3-year safety and efficacy of treating all-comer patients with 3 contemporary drug-eluting stents (DES).
The BIO-RESORT (Comparison of Biodegradable Polymer and Durable Polymer Drug-Eluting Stents in an All Comers Population) (TWENTE III) randomized trial (NCT01674803) found similar 1-year safety and efficacy for the 2 biodegradable-polymer DES (i.e., ultrathin-strut cobalt-chromium Orsiro sirolimus-eluting stent [SES] and very-thin-strut platinum-chromium Synergy everolimus-eluting stent) compared with the durable-polymer thin-strut cobalt-chromium Resolute Integrity zotarolimus-eluting stent (ZES). Two-year follow-up suggested that the SES might reduce repeat revascularizations beyond 1 year compared with the ZES.
A total of 3,514 all-comer patients were treated at 4 centers for coronary intervention. The main clinical endpoint, target vessel failure, was a composite of safety (cardiac death or target vessel-related myocardial infarction) and efficacy (target vessel revascularization). Secondary endpoints included the individual components of target vessel failure and stent thrombosis.
Three-year follow-up data were available for 3,393 of 3,514 patients (96.6%). Target vessel failure occurred in 8.5% with SES and 10.0% with ZES (plog rank = 0.22) and in 8.8% with everolimus-eluting stents (vs. ZES, plog rank = 0.32). Rates of cardiac death, target vessel myocardial infarction, and target vessel revascularization were similar between stent groups. Landmark analyses found no statistically significant between-stent difference in repeat revascularization between 1 and 3 years. Definite or probable stent thrombosis rates were low (SES, 1.1%; everolimus-eluting stent, 1.1%; ZES, 0.9%) and similar with all 3 DES.
Despite substantial differences in stent backbone and polymer coating, all 3 DES showed favorable 3-year safety and efficacy in all comers, without significant between-stent differences. Further follow-up is required to definitely answer the question of whether one stent might improve clinical outcomes at a later stage.
Buiten RA
,Ploumen EH
,Zocca P
,Doggen CJM
,Danse PW
,Schotborgh CE
,Scholte M
,van Houwelingen KG
,Stoel MG
,Hartmann M
,Tjon Joe Gin RM
,Somi S
,Linssen GCM
,Kok MM
,von Birgelen C
... -
《-》
-
Final 5-year report of BIONYX comparing the thin-composite wire-strut zotarolimus-eluting stent versus ultrathin-strut sirolimus-eluting stent.
The BIONYX randomized trial is the first study to evaluate the Resolute Onyx durable polymer-coated zotarolimus-eluting stent (ZES) in all-comers. Furthermore, it is the first trial to assess safety and efficacy of this stent versus the Orsiro biodegradable-polymer sirolimus-eluting stent (SES) in all-comers, paying particular attention to patients with diabetes. It has previously shown promising results until 3 years of follow-up.
We aimed to assess long-term clinical outcome after percutaneous coronary intervention (PCI) with Onyx ZES versus Orsiro SES at 5-year follow-up.
The main composite endpoint was target vessel failure (TVF): cardiac death, target vessel myocardial infarction, or target vessel revascularization. Time to primary and secondary endpoints was assessed using Kaplan-Meier methods, applying the log-rank test for between-group comparison.
Follow-up was available in 2414/2488 (97.0%) patients. After 5 years, TVF showed no significant difference between Onyx ZES and Orsiro SES (12.7% vs. 13.7%, hazard ratio [HR] 0.94, 95% confidence interval [CI] [0.75-1.17], plog-rank = 0.55). Landmark analysis between 3- and 5-year follow-up found a lower target lesion revascularization rate for Onyx ZES (1.1% vs. 2.4%, HR 0.47, 95% CI [0.24-0.93], plog-rank = 0.026). A prespecified subgroup analysis showed no significant between-stent difference in clinical outcome among patients with diabetes. After treatment with Onyx ZES, patients aged ≥75 years had significantly lower rates of TVF (13.8% vs. 21.9%, HR 0.60, 95% CI [0.39-0.93], plog-rank = 0.023).
The final 5-year analysis of the randomized BIONYX trial showed favorable and similar long-term outcomes of safety and efficacy for Onyx ZES and Orsiro SES in both all-comers and patients with diabetes.
van Vliet D
,Ploumen EH
,Pinxterhuis TH
,Buiten RA
,Aminian A
,Schotborgh CE
,Danse PW
,Roguin A
,Anthonio RL
,Benit E
,Zocca P
,Doggen CJM
,von Birgelen C
... -
《-》
-
Thin composite wire strut, durable polymer-coated (Resolute Onyx) versus ultrathin cobalt-chromium strut, bioresorbable polymer-coated (Orsiro) drug-eluting stents in allcomers with coronary artery disease (BIONYX): an international, single-blind, randomi
During the past decade, many patients had zotarolimus-eluting stents implanted, which had circular shape cobalt-chromium struts with limited radiographic visibility. The Resolute Onyx stent was developed to improve visibility while reducing strut thickness, which was achieved by using a novel composite wire with a dense platinum-iridium core and an outer cobalt-chromium layer. We did the first randomised clinical trial to assess the safety and efficacy of this often-used stent compared with the Orsiro stent, which consists of ultrathin cobalt-chromium struts.
We did an investigator-initiated, assessor-blinded and patient-blinded, randomised non-inferiority trial in an allcomers population at seven independently monitored centres in Belgium, Israel, and the Netherlands. Eligible participants were aged 18 years or older and required percutaneous coronary intervention with drug-eluting stents. After guide wire passage with or without predilation, members of the catheterisation laboratory team used web-based computer-generated allocation sequences to randomly assign patients (1:1) to either the Resolute Onyx or the Orsiro stent. Randomisation was stratified by sex and diabetes status. Patients and assessors were masked to allocated stents, but treating clinicians were not. The primary endpoint was target vessel failure at 1 year, a composite of cardiac death, target-vessel-related myocardial infarction, and target vessel revascularisation, and was assessed by intention to treat (non-inferiority margin 2·5%) on the basis of outcomes adjudicated by an independent event committee. This trial is registered with ClinicalTrials.gov, number NCT02508714.
Between Oct 7, 2015, and Dec 23, 2016, 2516 patients were enrolled, 2488 of whom were included in the intention-to-treat analysis (28 withdrawals or screening failures). 1243 participants were assigned to the Resolute Onyx group, and 1245 to the Orsiro group. Overall, 1765 (70·9%) participants presented with acute coronary syndromes and 1275 (51·2%) had myocardial infarctions. 1-year follow-up was available for 2478 (99·6%) patients. The primary endpoint was met by 55 (4·5%) patients in the Resolute Onyx group and 58 (4·7%) in the Orsiro group. Non-inferiority of Resolute Onyx to Orsiro was thus established (absolute risk difference -0·2% [95% CI -1·9 to 1·4]; upper limit of the one-sided 95% CI 1·1%; pnon-inferiority=0·0005). Definite or probable stent thrombosis occurred in one (0·1%) participant in the Resolute Onyx group and nine (0·7%) in the Orsiro group (hazard ratio 0·11 [95% CI 0·01-0·87]; p=0·0112).
The Resolute Onyx stent was non-inferior to Orsiro for a combined safety and efficacy endpoint at 1-year follow-up in allcomers. The low event rate in both groups suggests that both stents are safe, and the very low rate of stent thrombosis in the Resolute Onyx group warrants further clinical investigation.
Biotronik and Medtronic.
von Birgelen C
,Zocca P
,Buiten RA
,Jessurun GAJ
,Schotborgh CE
,Roguin A
,Danse PW
,Benit E
,Aminian A
,van Houwelingen KG
,Anthonio RL
,Stoel MG
,Somi S
,Hartmann M
,Linssen GCM
,Doggen CJM
,Kok MM
... -
《-》
-
5-Year Outcome Following Randomized Treatment of All-Comers With Zotarolimus-Eluting Resolute Integrity and Everolimus-Eluting PROMUS Element Coronary Stents: Final Report of the DUTCH PEERS (TWENTE II) Trial.
The study sought to evaluate for the first time the 5-year outcomes after treating an all-comers population with newer-generation cobalt chromium-based Resolute Integrity zotarolimus-eluting stents (ZES) (Medtronic, Santa Rosa, California) versus platinum chromium-based PROMUS Element everolimus eluting stents (EES) (Boston Scientific, Natick, Massachusetts).
The DUTCH PEERS (TWENTE II) (DUrable polymer-based sTent CHallenge of Promus ElemEnt versus ReSolute integrity: TWENTE II) trial is a randomized, multicenter, single-blinded, investigator-initiated all-comers trial that found at its main analysis similar 1-year safety and efficacy for both drug-eluting stents. It is the first randomized trial ever to investigate the Resolute Integrity ZES and the first trial to compare both devices.
In total, 1,811 patients were 1:1 randomized to ZES versus EES. We performed a pre-specified assessment of the 5-year clinical outcomes in terms of safety and efficacy. The main endpoint target vessel failure (TVF) is a composite of cardiac death, target vessel-related myocardial infarction, or target vessel revascularization. Secondary endpoints included the individual components of TVF, and stent thrombosis. The study was independently monitored, and adverse clinical events were independently adjudicated.
Five-year clinical follow-up data was available in 1,798 (99.3%) patients. The ZES and EES groups showed favorable outcomes, with similar 5-year incidence of TVF (13.2% vs. 14.2%; plog-rank = 0.62) and its individual components: cardiac death (4.5% vs. 4.9%; plog-rank = 0.69), target vessel-related myocardial infarction (3.1% vs. 2.6%; plog-rank = 0.47), and target vessel revascularization (7.6% vs. 8.6%; plog-rank = 0.46). The 5-year incidence of definite or probable stent thrombosis was similar (1.5% vs. 1.3%; plog-rank = 0.83).
At 5-year follow-up, the Resolute Integrity ZES and PROMUS Element EES showed similar and sustained results in terms of safety and efficacy for treating a broad population of all-comers.
Zocca P
,Kok MM
,Tandjung K
,Danse PW
,Jessurun GAJ
,Hautvast RWM
,van Houwelingen KG
,Stoel MG
,Schramm AR
,Tjon Joe Gin RM
,de Man FHAF
,Hartmann M
,Louwerenburg JHW
,Linssen GCM
,Löwik MM
,Doggen CJM
,von Birgelen C
... -
《-》