Outcome After Myocardial Infarction Treated With Resolute Integrity and Promus Element Stents: Insights From the DUTCH PEERS (TWENTE II) Randomized Trial.
In acute myocardial infarction (MI), novel highly deliverable drug-eluting stents (DES) may be particularly valuable as their flexible stent designs might reduce device-induced traumas to culprit lesions. The aim of the study was to assess the safety and efficacy of percutaneous coronary interventions with 2 novel durable polymer-coated DES in patients with acute MI.
The prospective, randomized DUTCH PEERS (TWENTE II) multicenter trial compares Resolute Integrity and Promus Element stents in 1811 all-comer patients, of whom 817 (45.1%) were treated for ST-segment elevation MI or non-ST-segment elevation MI and the 2-year outcome is available in 99.9%. The primary clinical endpoint is target vessel failure (TVF), a composite of cardiac death, target vessel related MI, or target vessel revascularization.
Of all 817 patients treated for acute MI, 421 (51.5%) were treated with Resolute Integrity and 396 (48.5%) with Promus Element stents. At the 2-year follow-up, the rates of TVF (7.4% vs 6.1%; P = .45), target lesion revascularization (3.1% vs 2.8%; P = .79), and definite stent thrombosis (1.0% vs 0.5%; P = .69) were low for both stent groups. Consistent with these findings in all patients with acute MI, outcomes for the 2 DES were favorable and similar in both, with 370 patients with ST-segment elevation MI (TVF, 5.1% vs 4.9%; P = .81) and 447 patients with non-ST-segment elevation MI (TVF, 9.0% vs 7.5%; P = .56).
Resolute Integrity and Promus Element stents were both safe and efficacious in treating patients with acute MI. The present 2-year follow-up data underline the safety of using these devices in this particular clinical setting.
van Houwelingen KG
,Lam MK
,Löwik MM
,Danse PW
,Tjon Joe Gin RM
,Jessurun GA
,Anthonio RL
,Sen H
,Linssen GCM
,IJzerman MJ
,Doggen CJM
,von Birgelen C
... -
《-》
5-Year Outcome Following Randomized Treatment of All-Comers With Zotarolimus-Eluting Resolute Integrity and Everolimus-Eluting PROMUS Element Coronary Stents: Final Report of the DUTCH PEERS (TWENTE II) Trial.
The study sought to evaluate for the first time the 5-year outcomes after treating an all-comers population with newer-generation cobalt chromium-based Resolute Integrity zotarolimus-eluting stents (ZES) (Medtronic, Santa Rosa, California) versus platinum chromium-based PROMUS Element everolimus eluting stents (EES) (Boston Scientific, Natick, Massachusetts).
The DUTCH PEERS (TWENTE II) (DUrable polymer-based sTent CHallenge of Promus ElemEnt versus ReSolute integrity: TWENTE II) trial is a randomized, multicenter, single-blinded, investigator-initiated all-comers trial that found at its main analysis similar 1-year safety and efficacy for both drug-eluting stents. It is the first randomized trial ever to investigate the Resolute Integrity ZES and the first trial to compare both devices.
In total, 1,811 patients were 1:1 randomized to ZES versus EES. We performed a pre-specified assessment of the 5-year clinical outcomes in terms of safety and efficacy. The main endpoint target vessel failure (TVF) is a composite of cardiac death, target vessel-related myocardial infarction, or target vessel revascularization. Secondary endpoints included the individual components of TVF, and stent thrombosis. The study was independently monitored, and adverse clinical events were independently adjudicated.
Five-year clinical follow-up data was available in 1,798 (99.3%) patients. The ZES and EES groups showed favorable outcomes, with similar 5-year incidence of TVF (13.2% vs. 14.2%; plog-rank = 0.62) and its individual components: cardiac death (4.5% vs. 4.9%; plog-rank = 0.69), target vessel-related myocardial infarction (3.1% vs. 2.6%; plog-rank = 0.47), and target vessel revascularization (7.6% vs. 8.6%; plog-rank = 0.46). The 5-year incidence of definite or probable stent thrombosis was similar (1.5% vs. 1.3%; plog-rank = 0.83).
At 5-year follow-up, the Resolute Integrity ZES and PROMUS Element EES showed similar and sustained results in terms of safety and efficacy for treating a broad population of all-comers.
Zocca P
,Kok MM
,Tandjung K
,Danse PW
,Jessurun GAJ
,Hautvast RWM
,van Houwelingen KG
,Stoel MG
,Schramm AR
,Tjon Joe Gin RM
,de Man FHAF
,Hartmann M
,Louwerenburg JHW
,Linssen GCM
,Löwik MM
,Doggen CJM
,von Birgelen C
... -
《-》
Clinical Events and Patient-Reported Chest Pain in All-Comers Treated With Resolute Integrity and Promus Element Stents: 2-Year Follow-Up of the DUTCH PEERS (DUrable Polymer-Based STent CHallenge of Promus ElemEnt Versus ReSolute Integrity) Randomized Tri
This study assessed clinical events and patient-reported chest pain 2 years after treatment of all-comers with Resolute Integrity zotarolimus-eluting stents (Medtronic Vascular, Santa Rosa, California) and Promus Element everolimus-eluting stents (Boston Scientific, Natick, Massachusetts).
For both drug-eluting stents (DES), no all-comer outcome data from >12 months of follow-up have been published. Although there is increasing interest in patient-reported chest pain following stenting, data with novel DES are scarce.
The DUTCH PEERS multicenter trial (TWENTE II) (DUrable Polymer-Based STent CHallenge of Promus ElemEnt Versus ReSolute Integrity) Randomized Trial [TWENTE II]) randomized 1,811 all-comer patients to treatment with 1 type of DES. Monitoring and event adjudication were performed by independent contract research organizations.
The 2-year follow-up of 1,810 patients (99.9%) was available. The primary composite endpoint target vessel failure occurred in 8.6% and 7.8% of patients treated with zotarolimus- and everolimus-eluting stents, respectively (p = 0.55). Rates of components of target vessel failure were: cardiac death (2.4% vs. 1.9%, p = 0.42); target vessel-related myocardial infarction (2.4% vs. 1.8%, p = 0.33); clinically-indicated target vessel revascularization (4.6% vs. 4.9%, p = 0.83). At 1- and 2-year follow-up, >80% of patients were free from chest pain (no between-stent difference). In addition, >87% of patients were either free from chest pain or experienced pain only at maximal physical exertion, but not during normal daily activities. Patients with chest pain after 12 months at no more than moderate physical effort had a higher risk of target vessel revascularization during the following year (hazard ratio: 1.89 [95% confidence interval: 1.05 to 3.39], p = 0.03).
During the second year of follow-up, the incidence of adverse clinical endpoints remained similar and low for both DES. The vast majority of patients were free from chest pain.
Sen H
,Lam MK
,Löwik MM
,Danse PW
,Jessurun GA
,van Houwelingen KG
,Anthonio RL
,Tjon Joe Gin RM
,Hautvast RW
,Louwerenburg JH
,de Man FH
,Stoel MG
,van der Heijden LC
,Linssen GC
,IJzerman MJ
,Tandjung K
,Doggen CJ
,von Birgelen C
... -
《-》
Third-generation zotarolimus-eluting and everolimus-eluting stents in all-comer patients requiring a percutaneous coronary intervention (DUTCH PEERS): a randomised, single-blind, multicentre, non-inferiority trial.
Third-generation, permanent-polymer-based drug-eluting stents with novel, flexible designs might be more easily delivered than previous generations of stents in complex coronary lesions, but might be less longitudinally stable. We aimed to assess the safety and efficacy in all-comer patients of two third-generation stents that are often used clinically, but that have not yet been compared, and one of which has not previously been assessed in a randomised trial.
In this investigator-initiated, single-blind, multicentre, randomised, two-arm, non-inferiority trial, patients aged 18 years and older who required a percutaneous coronary intervention with implantation of a drug-eluting stent were recruited from four study sites in the Netherlands. We randomly assigned patients by independently managed computer-generated allocation sequences in a 1:1 ratio to receive either cobalt-chromium-based zotarolimus-eluting stents (Resolute Integrity, Medtronic, Santa Rosa, CA, USA) or platinum-chromium-based everolimus-eluting stents (Promus Element, Boston Scientific, Natick, MA, USA). Patients and analysts were masked to the allocated stent, but treating clinicians were not. The primary endpoint of target-vessel failure was a composite of safety (cardiac death or target-vessel-related myocardial infarction) and efficacy (target-vessel revascularisation) at 12 months, analysed by intention to treat (with a non-inferiority margin of 3·6%). This trial is registered with ClinicalTrials.gov, number NCT01331707.
Between Nov 25, 2010, and May 24, 2012, 1811 eligible all-comer patients, with 2371 target lesions, were enrolled in the study. 370 (20%) patients presented with ST-elevation myocardial infarction and 447 (25%) with non-ST-elevation myocardial infarction. 906 patients were assigned to receive zotarolimus-eluting stents and 905 to receive everolimus-eluting stents. Ease of stent delivery was shown by very low numbers of patients requiring treatment other than their assigned study treatment (six [1%] in the zotarolimus-eluting stent group vs five [1%] in the everolimus-eluting stent group; p=0·22). 12-month follow-up results were available for 1810 patients (one patient in the zotarolimus-eluting stent group withdrew consent). The primary endpoint was met by 55 (6%) of 905 patients in the zotarolimus-eluting stent group and 47 (5%) of 905 in the everolimus-eluting stent group. The zotarolimus-eluting stent was non-inferior to the everolimus-eluting stent (absolute risk difference 0·88%, 95% CI -1·24% to 3·01%; upper limit of one-sided 95% CI 2·69%; non-inferiority p=0·006). We noted no significant between-group differences in individual components of the primary endpoint. Definite stent thrombosis occurred in three (0·3%) patients in the zotarolimus-eluting stent group and six (0·7%) patients in the everolimus-eluting stent group (p=0·34). Longitudinal stent deformation was seen only in the everolimus-eluting stent group (nine [1·0%] of 905 vs 0 of 906, p=0·002; nine of 1591 [0·6%] everolimus-eluting stents implanted became deformed), but was not associated with any adverse events.
Both stents were similarly efficacious and safe, and provided excellent clinical outcomes, especially in view of the large number of patients who presented with acute myocardial infarctions.
Boston Scientific, Medtronic.
von Birgelen C
,Sen H
,Lam MK
,Danse PW
,Jessurun GA
,Hautvast RW
,van Houwelingen GK
,Schramm AR
,Gin RM
,Louwerenburg JW
,de Man FH
,Stoel MG
,Löwik MM
,Linssen GC
,Saïd SA
,Nienhuis MB
,Verhorst PM
,Basalus MW
,Doggen CJ
,Tandjung K
... -
《-》