Characterization of a microDiamond detector in high-dose-per-pulse electron beams for intra operative radiation therapy.
To characterize a synthetic diamond dosimeter (PTW Freiburg microDiamond 60019) in high dose-per-pulse electron beams produced by an Intra Operative Radiation Therapy (IORT) dedicated accelerator.
The dosimetric properties of the microDiamond were assessed under 6, 8 and 9 MeV electron beams by a NOVAC11 mobile accelerator (Sordina IORT Technologies S.p.A.). The characterization was carried out with dose-per-pulse ranging from 26 to 105 mGy per pulse. The microDiamond performance was compared with an Advanced Markus ionization chamber and a PTW silicon diode E in terms of dose linearity, percentage depth dose (PDD) curves, beam profiles and output factors.
A good linearity of the microDiamond response was verified in the dose range from 0.2 Gy to 28 Gy. A sensitivity of 1.29 nC/Gy was measured under IORT electron beams, resulting within 1% with respect to the one obtained in reference condition under (60)Co gamma irradiation. PDD measurements were found in agreement with the ones by the reference dosimeters, with differences in R50 values below 0.3 mm. Profile measurements evidenced a high spatial resolution of the microDiamond, slightly worse than the one of the silicon diode. The penumbra widths measured by the microDiamond resulted approximately 0.5 mm larger than the ones by the Silicon diode. Output factors measured by the microDiamond were found within 2% with those obtained by the Advanced Markus down to 3 cm diameter field sizes.
The microDiamond dosimeter was demonstrated to be suitable for precise dosimetry in IORT applications under high dose-per-pulse conditions.
Di Venanzio C
,Marinelli M
,Tonnetti A
,Verona-Rinati G
,Falco MD
,Pimpinella M
,Ciccotelli A
,De Stefano S
,Felici G
,Marangoni F
... -
《-》
Output factors of ionization chambers and solid state detectors for mobile intraoperative radiotherapy (IORT) accelerator electron beams.
The electron energy characteristics of mobile intraoperative radiotherapy (IORT) accelerator LIAC® differ from commonly used linear accelerators, thus some of the frequently used detectors can give less accurate results. The aim of this study is to evaluate the output factors (OFs) of several ionization chambers (IC) and solid state detectors (SS) for electron beam energies generated by LIAC® and compare with the output factor of Monte Carlo model (MC) in order to determine the adequate detectors for LIAC® .
The OFs were measured for 6, 8, 10, and 12 MeV electron energies with PTW 23343 Markus, PTW 34045 Advanced Markus, PTW 34001 Roos, IBA PPC05, IBA PPC40, IBA NACP-02, PTW 31010 Semiflex, PTW 31021 Semiflex 3D, PTW 31014 Pinpoint, PTW 60017 Diode E, PTW 60018 Diode SRS, SNC Diode EDGE, and PTW 60019 micro Diamond detectors. Ion recombination factors (ksat ) of IC were measured for all applicator sizes and OFs were corrected according to ksat . The measured OFs were compared with Monte Carlo output factors (OFMC ).
The measured OFs of IBA PPC05, PTW Advanced Markus, PTW Pinpoint, PTW microDiamond, and PTW Diode E detectors are in good agreement with OFMC . The maximum deviations of IBA PPC05 OFs to OFMC are -1.6%, +1.5%, +1.5%, and +2.0%; for PTW Advanced Markus +1.0%, +1.5%, +2.0%, and +2.0%; for PTW Pinpoint +2.0%, +1.6%, +4.0%, and +2.0%; for PTW microDiamond -1.6%, +2%, +1.1%, and +1.0%; and for PTW Diode E -+1.7%, +1.7%, +1.3%, and +2.5% for 6, 8, 10, and 12 MeV, respectively. PTW Roos, PTW Markus, IBA PPC40, PTW Semiflex, PTW Semiflex 3D, SNC Diode Edge measured OFs with a maximum deviation of +5.6%, +4.5%, +5.6%, +8.1%, +4.8%, and +9.6% with respect to OFMC , while PTW Diode SRS and IBA NACP-02 were the least accurate (with highest deviations -37.1% and -18.0%, respectively).
The OFs results of solid state detectors PTW microDiamond and PTW Diode E as well as the ICs with small electrode spacing distance such as IBA PPC05, PTW Advanced Markus and PTW Pinpoint are in excellent agreement with OFMC . The measurements of the other detectors evaluated in this study are less accurate, thus they should be used with caution. Particularly, PTW Diode SRS and IBA NACP-02 are not suitable and their use should be avoided in relative dosimetry measurements under high dose per pulsed (DPP) electron beams.
Güngör G
,Aydın G
,Mustafayev TZ
,Özyar E
... -
《Journal of Applied Clinical Medical Physics》
Use of PTW-microDiamond for relative dosimetry of unflattened photon beams.
The increasing interest in SBRT treatments encourages the use of flattening filter free (FFF) beams. Aim of this work was to evaluate the performance of the PTW60019 microDiamond detector under 6MV and 10MVFFF beams delivered with the EDGE accelerator (Varian Medical System, Palo Alto, USA). A flattened 6MV beam was also considered for comparison.
Short term stability, dose linearity and dose rate dependence were evaluated. Dose per pulse dependence was investigated in the range 0.2-2.2mGy/pulse. MicroDiamond profiles and output factors (OFs) were compared to those obtained with other detectors for field sizes ranging from 40×40cm2 to 0.6×0.6cm2. In small fields, volume averaging effects were evaluated and the relevant correction factors were applied for each detector.
MicroDiamond short term stability, dose linearity and dependence on monitor unit rate were less than 0.8% for all energies. Response variations with dose per pulse were found within 1.8%. MicroDiamond output factors (OF) values differed from those measured with the reference ion-chamber for less than 1% up to 40×40cm2 fields where silicon diodes overestimate the dose of ≈3%. For small fields (<3×3cm2) microDiamond and the unshielded silicon diode were in good agreement.
MicroDiamond showed optimal characteristics for relative dosimetry even under high dose rate beams. The effects due to dose per pulse dependence up to 2.2mGy/pulse are negligible. Compared to other detectors, microDiamond provides accurate OF measurements in the whole range of field sizes. For fields <1cm correction factors accounting for fluence perturbation and volume averaging could be required.