Minimizing ultraviolet noise due to mis-matches between detector flow cell and post column mobile phase temperatures in supercritical fluid chromatography: effect of flow cell design.
摘要:
A mis-match between the post-column mobile phase temperature and the UV detector flow cell temperature can cause significant UV noise in supercritical fluid chromatography (SFC). Deviations as little as 5 °C can increase noise as much as 5 times, making the detector unsuited for trace analysis. Two approaches were used to minimize this noise. When a flow cell was in direct thermal contact (metal on metal) with the detector optical bench, the mobile phase temperature was actively controlled to the measured flow cell temperature, by using one of the heat exchangers (HX) in the column compartment. However, with some older, but still widely used flow cell designs, this required repeated, hourly monitoring of the flow cell temperature and repeated manual adjustment of the heat exchanger temperature, due to thermal drift. Flow cell design had a strong influence on susceptibility to this thermally induced noise. Thermally insulating the flow cell from the optical bench made some cells much less susceptible to such thermally induced noise. Five different flow cells, some insulated, some un-insulated, were evaluated. Most had a truncated conical flow path, but one had a cylindrical flow path. Using either approach, the ASTM noise, with a 10mm, 13 μL conical flow cell, could be optimized to ≈0.007 mAU at 2.5 Hz, in SFC, which is very near the 0.006 mAU manufacturer's specification for HPLC. The insulated version of this flow cell required far less optimization, compared to the un-insulated version. At 150 bar, an experimental 3mm, 2 μL flow cell, with only one side insulated, yielded noise slightly too high (≈0.16-0.18 mAU) for trace analysis, at 80 Hz. However, at 200 bar, noise at 80 Hz was <0.06 mAU, which should allow quantification of a 1 mAU tall trace component with a signal to noise ratio (S/N) >10. Even partially un-insulated, this flow cell design was much less susceptible to thermally induced noise. Further insulating this flow cell design failed to improve performance.
收起
展开
DOI:
10.1016/j.chroma.2014.08.067
被引量:
年份:
1970


通过 文献互助 平台发起求助,成功后即可免费获取论文全文。
求助方法1:
知识发现用户
每天可免费求助50篇
求助方法1:
关注微信公众号
每天可免费求助2篇
求助方法2:
完成求助需要支付5财富值
您目前有 1000 财富值
相似文献(119)
参考文献(0)
引证文献(0)
来源期刊
影响因子:暂无数据
JCR分区: 暂无
中科院分区:暂无