Inspection of the human stomach using remote-controlled capsule endoscopy: a feasibility study in healthy volunteers (with videos).
Remote control of capsule endoscopes might allow reliable inspection of the human stomach.
To assess the safety and efficacy of manipulation of a modified capsule endoscope with magnetic material (magnetic maneuverable capsule [MMC]) in the human stomach by using a handheld external magnet.
Open clinical trial.
Academic hospital.
Ten healthy volunteers.
Subjects swallowed the MMC and sherbet powder for gastric distention. An external magnetic paddle (EMP-2) was used to manipulate the MMC within the stomach. MMC responsiveness was evaluated on a screen showing the MMC film in real time.
Safety and tolerability (questionnaire), gastric residence time of the MMC, its responsiveness to the EMP-2, area of gastric mucosa visualized.
There were no adverse events. The MMC was always clearly attracted by the EMP-2 and responded to its movements. It remained in the stomach for 39 ± 24 minutes. In 7 subjects, both the cardia and the pylorus were inspected and 75% or more of the gastric mucosa was visualized (≥50% in all of the remaining subjects). A learning curve was clearly recognizable (identification of MMC localization, intended movements).
Small amounts of fluid blocked the view of apical parts of the fundus; gastric distention was not sufficient to flatten all gastric folds.
Remote control of the MMC in the stomach of healthy volunteers using a handheld magnet is safe and feasible. Responsiveness of the MMC was excellent, and visualization of the gastric mucosa was good, although not yet complete, in the majority of subjects. The system appeared to be clinically valuable and should be developed further. (
DE/CA05/2009031008.).
Keller J
,Fibbe C
,Volke F
,Gerber J
,Mosse AC
,Reimann-Zawadzki M
,Rabinovitz E
,Layer P
,Schmitt D
,Andresen V
,Rosien U
,Swain P
... -
《-》
Accuracy of Magnetically Controlled Capsule Endoscopy, Compared With Conventional Gastroscopy, in Detection of Gastric Diseases.
Diseases of the stomach, including gastric cancer and peptic ulcer, are the most common digestive diseases. It is impossible to visualize the entire stomach with the passive capsule currently used in practice because of the large size of the gastric cavity. A magnetically controlled capsule endoscopy (MCE) system has been designed to explore the stomach. We performed a prospective study to compare the accuracy of detection of gastric focal lesions by MCE vs conventional gastroscopy (the standard method).
We performed a multicenter blinded study comparing MCE with conventional gastroscopy in 350 patients (mean age, 46.6 y), with upper abdominal complaints scheduled to undergo gastroscopy at a tertiary center in China from August 2014 through December 2014. All patients underwent MCE, followed by conventional gastroscopy 2 hours later, without sedation. We calculated the sensitivity, specificity, positive predictive value, and negative predictive value of detection of gastric focal lesions by MCE, using gastroscopy as the standard.
MCE detected gastric focal lesions in the whole stomach with 90.4% sensitivity (95% confidence interval [CI], 84.7%-96.1%), 94.7% specificity (95% CI, 91.9%-97.5%), a positive predictive value of 87.9% (95% CI, 81.7%-94.0%), a negative predictive value of 95.9% (95% CI, 93.4%-98.4%), and 93.4% accuracy (95% CI, 90.83%-96.02%). MCE detected focal lesions in the upper stomach (cardia, fundus, and body) with 90.2% sensitivity (95% CI, 82.0%-98.4%) and 96.7% specificity (95% CI, 94.4%-98.9%). MCE detected focal lesions in the lower stomach (angulus, antrum, and pylorus) with 90.6% sensitivity (95% CI, 82.7%-98.4%) and 97.9% specificity (95% CI, 96.1%-99.7%). MCE detected 1 advanced gastric carcinoma, 2 malignant lymphomas, and 1 early stage gastric tumor. MCE did not miss any lesions of significance (including tumors or large ulcers). Among the 350 patients, 5 reported 9 adverse events (1.4%) and 335 preferred MCE over gastroscopy (95.7%).
MCE detects focal lesions in the upper and lower stomach with comparable accuracy with conventional gastroscopy. MCE is preferred by almost all patients, compared with gastroscopy, and can be used to screen gastric diseases without sedation. Clinicaltrials.gov number: NCT02219529.