Feasibility, diagnostic accuracy, and effectiveness of decentralised use of the Xpert MTB/RIF test for diagnosis of tuberculosis and multidrug resistance: a multicentre implementation study.
The Xpert MTB/RIF test (Cepheid, Sunnyvale, CA, USA) can detect tuberculosis and its multidrug-resistant form with very high sensitivity and specificity in controlled studies, but no performance data exist from district and subdistrict health facilities in tuberculosis-endemic countries. We aimed to assess operational feasibility, accuracy, and effectiveness of implementation in such settings.
We assessed adults (≥18 years) with suspected tuberculosis or multidrug-resistant tuberculosis consecutively presenting with cough lasting at least 2 weeks to urban health centres in South Africa, Peru, and India, drug-resistance screening facilities in Azerbaijan and the Philippines, and an emergency room in Uganda. Patients were excluded from the main analyses if their second sputum sample was collected more than 1 week after the first sample, or if no valid reference standard or MTB/RIF test was available. We compared one-off direct MTB/RIF testing in nine microscopy laboratories adjacent to study sites with 2-3 sputum smears and 1-3 cultures, dependent on site, and drug-susceptibility testing. We assessed indicators of robustness including indeterminate rate and between-site performance, and compared time to detection, reporting, and treatment, and patient dropouts for the techniques used.
We enrolled 6648 participants between Aug 11, 2009, and June 26, 2010. One-off MTB/RIF testing detected 933 (90·3%) of 1033 culture-confirmed cases of tuberculosis, compared with 699 (67·1%) of 1041 for microscopy. MTB/RIF test sensitivity was 76·9% in smear-negative, culture-positive patients (296 of 385 samples), and 99·0% specific (2846 of 2876 non-tuberculosis samples). MTB/RIF test sensitivity for rifampicin resistance was 94·4% (236 of 250) and specificity was 98·3% (796 of 810). Unlike microscopy, MTB/RIF test sensitivity was not significantly lower in patients with HIV co-infection. Median time to detection of tuberculosis for the MTB/RIF test was 0 days (IQR 0-1), compared with 1 day (0-1) for microscopy, 30 days (23-43) for solid culture, and 16 days (13-21) for liquid culture. Median time to detection of resistance was 20 days (10-26) for line-probe assay and 106 days (30-124) for conventional drug-susceptibility testing. Use of the MTB/RIF test reduced median time to treatment for smear-negative tuberculosis from 56 days (39-81) to 5 days (2-8). The indeterminate rate of MTB/RIF testing was 2·4% (126 of 5321 samples) compared with 4·6% (441 of 9690) for cultures.
The MTB/RIF test can effectively be used in low-resource settings to simplify patients' access to early and accurate diagnosis, thereby potentially decreasing morbidity associated with diagnostic delay, dropout and mistreatment.
Foundation for Innovative New Diagnostics, Bill & Melinda Gates Foundation, European and Developing Countries Clinical Trials Partnership (TA2007.40200.009), Wellcome Trust (085251/B/08/Z), and UK Department for International Development.
Boehme CC
,Nicol MP
,Nabeta P
,Michael JS
,Gotuzzo E
,Tahirli R
,Gler MT
,Blakemore R
,Worodria W
,Gray C
,Huang L
,Caceres T
,Mehdiyev R
,Raymond L
,Whitelaw A
,Sagadevan K
,Alexander H
,Albert H
,Cobelens F
,Cox H
,Alland D
,Perkins MD
... -
《-》
Xpert MTB/RIF Ultra for detection of Mycobacterium tuberculosis and rifampicin resistance: a prospective multicentre diagnostic accuracy study.
The Xpert MTB/RIF assay is an automated molecular test that has improved the detection of tuberculosis and rifampicin resistance, but its sensitivity is inadequate in patients with paucibacillary disease or HIV. Xpert MTB/RIF Ultra (Xpert Ultra) was developed to overcome this limitation. We compared the diagnostic performance of Xpert Ultra with that of Xpert for detection of tuberculosis and rifampicin resistance.
In this prospective, multicentre, diagnostic accuracy study, we recruited adults with pulmonary tuberculosis symptoms presenting at primary health-care centres and hospitals in eight countries (South Africa, Uganda, Kenya, India, China, Georgia, Belarus, and Brazil). Participants were allocated to the case detection group if no drugs had been taken for tuberculosis in the past 6 months or to the multidrug-resistance risk group if drugs for tuberculosis had been taken in the past 6 months, but drug resistance was suspected. Demographic information, medical history, chest imaging results, and HIV test results were recorded at enrolment, and each participant gave at least three sputum specimen on 2 separate days. Xpert and Xpert Ultra diagnostic performance in the same sputum specimen was compared with culture tests and drug susceptibility testing as reference standards. The primary objectives were to estimate and compare the sensitivity of Xpert Ultra test with that of Xpert for detection of smear-negative tuberculosis and rifampicin resistance and to estimate and compare Xpert Ultra and Xpert specificities for detection of rifampicin resistance. Study participants in the case detection group were included in all analyses, whereas participants in the multidrug-resistance risk group were only included in analyses of rifampicin-resistance detection.
Between Feb 18, and Dec 24, 2016, we enrolled 2368 participants for sputum sampling. 248 participants were excluded from the analysis, and 1753 participants were distributed to the case detection group (n=1439) and the multidrug-resistance risk group (n=314). Sensitivities of Xpert Ultra and Xpert were 63% and 46%, respectively, for the 137 participants with smear-negative and culture-positive sputum (difference of 17%, 95% CI 10 to 24); 90% and 77%, respectively, for the 115 HIV-positive participants with culture-positive sputum (13%, 6·4 to 21); and 88% and 83%, respectively, across all 462 participants with culture-positive sputum (5·4%, 3·3 to 8·0). Specificities of Xpert Ultra and Xpert for case detection were 96% and 98% (-2·7%, -3·9 to -1·7) overall, and 93% and 98% for patients with a history of tuberculosis. Xpert Ultra and Xpert performed similarly in detecting rifampicin resistance.
For tuberculosis case detection, sensitivity of Xpert Ultra was superior to that of Xpert in patients with paucibacillary disease and in patients with HIV. However, this increase in sensitivity came at the expense of a decrease in specificity.
Government of Netherlands, Government of Australia, Bill & Melinda Gates Foundation, Government of the UK, and the National Institute of Allergy and Infectious Diseases.
Dorman SE
,Schumacher SG
,Alland D
,Nabeta P
,Armstrong DT
,King B
,Hall SL
,Chakravorty S
,Cirillo DM
,Tukvadze N
,Bablishvili N
,Stevens W
,Scott L
,Rodrigues C
,Kazi MI
,Joloba M
,Nakiyingi L
,Nicol MP
,Ghebrekristos Y
,Anyango I
,Murithi W
,Dietze R
,Lyrio Peres R
,Skrahina A
,Auchynka V
,Chopra KK
,Hanif M
,Liu X
,Yuan X
,Boehme CC
,Ellner JJ
,Denkinger CM
,study team
... -
《-》
Xpert MTB/RIF and Xpert MTB/RIF Ultra assays for active tuberculosis and rifampicin resistance in children.
Every year, at least one million children become ill with tuberculosis and around 200,000 children die. Xpert MTB/RIF and Xpert Ultra are World Health Organization (WHO)-recommended rapid molecular tests that simultaneously detect tuberculosis and rifampicin resistance in adults and children with signs and symptoms of tuberculosis, at lower health system levels. To inform updated WHO guidelines on molecular assays, we performed a systematic review on the diagnostic accuracy of these tests in children presumed to have active tuberculosis.
Primary objectives • To determine the diagnostic accuracy of Xpert MTB/RIF and Xpert Ultra for (a) pulmonary tuberculosis in children presumed to have tuberculosis; (b) tuberculous meningitis in children presumed to have tuberculosis; (c) lymph node tuberculosis in children presumed to have tuberculosis; and (d) rifampicin resistance in children presumed to have tuberculosis - For tuberculosis detection, index tests were used as the initial test, replacing standard practice (i.e. smear microscopy or culture) - For detection of rifampicin resistance, index tests replaced culture-based drug susceptibility testing as the initial test Secondary objectives • To compare the accuracy of Xpert MTB/RIF and Xpert Ultra for each of the four target conditions • To investigate potential sources of heterogeneity in accuracy estimates - For tuberculosis detection, we considered age, disease severity, smear-test status, HIV status, clinical setting, specimen type, high tuberculosis burden, and high tuberculosis/HIV burden - For detection of rifampicin resistance, we considered multi-drug-resistant tuberculosis burden • To compare multiple Xpert MTB/RIF or Xpert Ultra results (repeated testing) with the initial Xpert MTB/RIF or Xpert Ultra result SEARCH METHODS: We searched the Cochrane Infectious Diseases Group Specialized Register, MEDLINE, Embase, Science Citation Index, the Cumulative Index to Nursing and Allied Health Literature (CINAHL), Scopus, the WHO International Clinical Trials Registry Platform, ClinicalTrials.gov, and the International Standard Randomized Controlled Trials Number (ISRCTN) Registry up to 29 April 2019, without language restrictions.
Randomized trials, cross-sectional trials, and cohort studies evaluating Xpert MTB/RIF or Xpert Ultra in HIV-positive and HIV-negative children younger than 15 years. Reference standards comprised culture or a composite reference standard for tuberculosis and drug susceptibility testing or MTBDRplus (molecular assay for detection of Mycobacterium tuberculosis and drug resistance) for rifampicin resistance. We included studies evaluating sputum, gastric aspirate, stool, nasopharyngeal or bronchial lavage specimens (pulmonary tuberculosis), cerebrospinal fluid (tuberculous meningitis), fine needle aspirates, or surgical biopsy tissue (lymph node tuberculosis).
Two review authors independently extracted data and assessed study quality using the Quality Assessment of Studies of Diagnostic Accuracy - Revised (QUADAS-2). For each target condition, we used the bivariate model to estimate pooled sensitivity and specificity with 95% confidence intervals (CIs). We stratified all analyses by type of reference standard. We assessed certainty of evidence using the GRADE approach.
For pulmonary tuberculosis, 299 data sets (68,544 participants) were available for analysis; for tuberculous meningitis, 10 data sets (423 participants) were available; for lymph node tuberculosis, 10 data sets (318 participants) were available; and for rifampicin resistance, 14 data sets (326 participants) were available. Thirty-nine studies (80%) took place in countries with high tuberculosis burden. Risk of bias was low except for the reference standard domain, for which risk of bias was unclear because many studies collected only one specimen for culture. Detection of pulmonary tuberculosis For sputum specimens, Xpert MTB/RIF pooled sensitivity (95% CI) and specificity (95% CI) verified by culture were 64.6% (55.3% to 72.9%) (23 studies, 493 participants; moderate-certainty evidence) and 99.0% (98.1% to 99.5%) (23 studies, 6119 participants; moderate-certainty evidence). For other specimen types (nasopharyngeal aspirate, 4 studies; gastric aspirate, 14 studies; stool, 11 studies), Xpert MTB/RIF pooled sensitivity ranged between 45.7% and 73.0%, and pooled specificity ranged between 98.1% and 99.6%. For sputum specimens, Xpert Ultra pooled sensitivity (95% CI) and specificity (95% CI) verified by culture were 72.8% (64.7% to 79.6%) (3 studies, 136 participants; low-certainty evidence) and 97.5% (95.8% to 98.5%) (3 studies, 551 participants; high-certainty evidence). For nasopharyngeal specimens, Xpert Ultra sensitivity (95% CI) and specificity (95% CI) were 45.7% (28.9% to 63.3%) and 97.5% (93.7% to 99.3%) (1 study, 195 participants). For all specimen types, Xpert MTB/RIF and Xpert Ultra sensitivity were lower against a composite reference standard than against culture. Detection of tuberculous meningitis For cerebrospinal fluid, Xpert MTB/RIF pooled sensitivity and specificity, verified by culture, were 54.0% (95% CI 27.8% to 78.2%) (6 studies, 28 participants; very low-certainty evidence) and 93.8% (95% CI 84.5% to 97.6%) (6 studies, 213 participants; low-certainty evidence). Detection of lymph node tuberculosis For lymph node aspirates or biopsies, Xpert MTB/RIF pooled sensitivity and specificity, verified by culture, were 90.4% (95% CI 55.7% to 98.6%) (6 studies, 68 participants; very low-certainty evidence) and 89.8% (95% CI 71.5% to 96.8%) (6 studies, 142 participants; low-certainty evidence). Detection of rifampicin resistance Xpert MTB/RIF pooled sensitivity and specificity were 90.0% (67.6% to 97.5%) (6 studies, 20 participants; low-certainty evidence) and 98.3% (87.7% to 99.8%) (6 studies, 203 participants; moderate-certainty evidence).
We found Xpert MTB/RIF sensitivity to vary by specimen type, with gastric aspirate specimens having the highest sensitivity followed by sputum and stool, and nasopharyngeal specimens the lowest; specificity in all specimens was > 98%. Compared with Xpert MTB/RIF, Xpert Ultra sensitivity in sputum was higher and specificity slightly lower. Xpert MTB/RIF was accurate for detection of rifampicin resistance. Xpert MTB/RIF was sensitive for diagnosing lymph node tuberculosis. For children with presumed tuberculous meningitis, treatment decisions should be based on the entirety of clinical information and treatment should not be withheld based solely on an Xpert MTB/RIF result. The small numbers of studies and participants, particularly for Xpert Ultra, limits our confidence in the precision of these estimates.
Kay AW
,González Fernández L
,Takwoingi Y
,Eisenhut M
,Detjen AK
,Steingart KR
,Mandalakas AM
... -
《Cochrane Database of Systematic Reviews》