Evaluation of the Endurant stent graft under instructions for use vs off-label conditions for endovascular aortic aneurysm repair.
This study evaluated the early and intermediate results of endovascular aortic aneurysm repair (EVAR) using the Endurant stent graft (Medtronic Cardiovascular, Santa Rosa, Calif) in patients treated according to device-specific instructions for use (IFU) for the proximal aortic neck compared with those obtained in patients treated in an off-label (OL) situation.
Between November 2007 and March 2010, 177 consecutive patients with abdominal aortic aneurysms (AAAs) were treated with the Endurant stent graft at our centers. The IFU for the Endurant stent graft included a proximal neck of 15 mm in length and <75° of angulation or 10 mm of neck length and <60° of angulation. The 121 patients (68.4%) operated on according to IFU were compared with 56 (31.6%) who underwent EVAR in OL circumstances to evaluate significant differences in demographics, intraoperative technical factors, and early (30 days) and intermediate outcomes (1 year).
Significantly more patients were aged >80 years in the OL group (37.5% vs 19%, P = .008), and they also had larger aneurysms (59 ± 10.6 vs 55.9 ± 10.8 mm, P = .05) and required a longer procedure time (69.3 ± 27.2 vs 60.8 ± 20.4 minutes, P = .02). At 30 days, the risk of type I endoleak was higher in the OL group (2 patients, 3.6% vs 0 in IFU), but this did not reach statistical significance (P = .09). The two groups were similar in rates of perioperative mortality, major morbidity, technical success, clinical success, complications, and reinterventions. At 1 year, there were no differences between the two groups in survival, freedom from any device-related reinterventions, and freedom from graft thrombosis. Estimated 1-year freedom from type I endoleak was 100% in the IFU group vs 93.3% in the OL group (P = .01).
In patients with both normal and complex anatomy of the proximal aortic neck, the Endurant stent graft obtained acceptable results, with no difference in survival, morbidity, or reinterventions. However, there was a greater risk of type I endoleak when OL indications were applied. Longer term follow-up is required to evaluate the effectiveness of this endograft in preventing late aneurysm-related complications.
Torsello G
,Troisi N
,Donas KP
,Austermann M
... -
《-》
One-year outcomes from an international study of the Ovation Abdominal Stent Graft System for endovascular aneurysm repair.
This study evaluated 1-year safety and effectiveness outcomes of the United States regulatory trial for the Ovation Abdominal Stent Graft System (TriVascular Inc, Santa Rosa, Calif) for endovascular repair of abdominal aortic aneurysms (AAAs).
This prospective, multicenter, single-arm trial was conducted at 36 sites in the United States, Germany, and Chile to evaluate the safety and effectiveness of the Ovation stent graft. From November 2009 to May 2011, 161 patients (88% males; mean age, 73 ± 8 years) with AAAs (mean diameter, 54 ± 9 mm) were treated with the Ovation stent graft. The main body is a modular two-docking limb device with a 14F outer diameter delivery system, active suprarenal fixation, and polymer-filled proximal rings that accommodate the aortic neck for seal. Main inclusion criteria included proximal aortic neck length ≥ 7 mm, inner neck diameter between 16 and 30 mm, distal iliac landing zones length ≥ 10 mm, and diameter between 8 and 20 mm. Patients were treated under a common protocol, including clinical and imaging follow-up at discharge, 30 days, 6 months, and annually through 5 years. A Clinical Events Committee adjudicated adverse events, an independent imaging core laboratory analyzed imaging, and a Data Safety and Monitoring Board provided study oversight. Complete 1-year follow-up data were available for this report.
The Ovation stent graft was implanted successfully in 161 patients (100%), including 69 (42.9%) by percutaneous access. General anesthesia was used in 106 patients (65.8%). Technical success was 100%, and mean procedure time was 110 minutes. Median procedural blood loss was 150 mL, and median hospital stay was 1 day. The 30-day major adverse event rate was 2.5%. At 1 year, AAA-related and all-cause mortality were 0.6% and 2.5%, respectively. Major adverse event and serious adverse event rates through 1 year were 6.2% and 38.5%, respectively. The 1-year treatment success rate was 99.3%. The imaging core laboratory reported no stent graft migration or type I, III, or IV endoleaks. At 1 year, type II endoleaks were identified in 34% of patients, and AAA enlargement was identified in one patient (0.7%). No AAA rupture or conversion to open surgery was reported. AAA-related secondary procedures were performed in 10 patients (6.2%) for 12 findings, including endoleak (six), aortic main body stenosis (three), and iliac limb stenosis or occlusion (three).
The 1-year results of the Ovation Abdominal Stent Graft System demonstrate excellent safety and effectiveness in treatment of patients with AAAs, particularly in patients with challenging anatomic characteristics, including short aortic necks and narrow iliac arteries. Longer-term follow-up is needed.
Mehta M
,Valdés FE
,Nolte T
,Mishkel GJ
,Jordan WD
,Gray B
,Eskandari MK
,Botti C
,A Pivotal Clinical Study to Evaluate the Safety and Effectiveness of the Ovation Abdominal Stent Graft System Investigators
... -
《-》
Propensity-matched cohort validates findings of the VALOR trial.
The Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms (VALOR) trial findings noted superior 30-day and 1-year outcomes of the Talent thoracic endograft (Medtronic Vascular, Santa Rosa, Calif) compared with surgical repair of descending thoracic aneurysms (DTAs). Data from 195 prospective thoracic endovascular aneurysm repair (TEVAR) patients treated with the Talent device and 189 retrospective controls undergoing open surgical repair (OSR) from three centers of excellence were included in the trial after completion of TEVAR enrollment and compared. Such comparisons are biased by baseline differences among TEVAR vs OSR, however, propensity score (PS) analysis can reduce bias and validate such comparisons.
Logistic regression was used to generate a PS (range, 0-1) to identify baseline characteristics more likely in TEVAR. The PS estimated the probability that any patient would undergo TEVAR (eg, a PS of 0.99 represents a 99% chance a patient belongs to TEVAR). PSs were then generated for all patients, and TEVAR and OSR patients were divided into tertiles based on the PS to reduce up to 80% of inherent bias. Outcomes from the middle tertile (T2), patients equally likely (midrange PS) to be in TEVAR or OSR and therefore best matched, were compared using regression analysis and were also compared with the outcomes in the overall trial group.
Correlates of membership in TEVAR were smaller aneurysm (P < .001), anticoagulants (P < .01), no previous abdominal aortic aneurysm (AAA) repair (P < .01), no peripheral vascular disease (P = .001), statin use (P = .002), aspirin use (P = .002), older age (P = .028), race (P = .007), male gender (P = .02), and heart failure (P = .035). T2 included 68 TEVAR (PS, 0.58 ± 0.2) and 67 OSR patients (PS, 0.46 ± 0.2). VALOR overall reported differences in aneurysm size (56 mm TEVAR vs 69 mm open) and prior AAA repair (19% TEVAR vs 37% open), and this adjusted to no differences in T2 patients. In the well-matched T2 cohort, TEVAR patients had similar 30-day mortality (0% vs 3% OSR; P = .2) and improved 1-year aneurysm-related mortality rates (0% TEVAR vs 8% OSR; P = .05) compared with the OSR patients. This finding was in concurrence with the VALOR trial reporting similar benefit in TEVAR patients. The all-cause 1-year mortality showed a favorable trend for TEVAR in the VALOR trial; however, in T2 patients, 1-year all-cause mortality was similar in both groups of patients (17% TEVAR vs 15% OSR; P = .8). Age (P = .01), history of cerebrovascular accident (P < .05), antiarrhythmia medication (P = .04), and renal disease (P < .03) independently predicted all-cause and aneurysm-related mortality by regression analysis.
PS analysis is an important tool for elimination of bias inherent when retrospective controls are used. Its application to VALOR validates the long-term benefit in aneurysm-related mortality conferred by TEVAR in patients undergoing endovascular DTA repair.
Patel VI
,Conrad MF
,Kwolek CJ
,Ouriel K
,Fairman RM
,Cambria RP
... -
《-》